Skip to main content
. 2021 Oct 16;20:34–39. doi: 10.1016/j.phro.2021.10.001

Table 4.

Gamma analysis results for all patients (n = 40), prostate cohort (n = 20) and greater pelvis cohort (n = 20) for plans on sCT with a derived curve REDsCT,mod applied.

3%/2 mm
2%/2 mm
1%/1 mm
Pass Rate (%) Av Gamma Pass Rate (%) Av Gamma Pass Rate (%) Av Gamma
All Patients 99.8 ± 0.5 0.14 ± 0.04 99.6 ± 0.9 0.2 ± 0.1 92.9 ± 5.7 0.4 ± 0.1
range 100.0–97.5 0.08–0.27 100.0–95.3 0.12–0.37 98.3–71.1 0.24–0.77
Prostate 100.0 ± 0.1 0.12 ± 0.02 99.9 ± 0.1 0.16 ± 0.03 96.0 ± 2.0 0.30 ± 0.08
range 100.0–99.8 0.08–0.16 100.0–99.5 0.12–0.25 98.3–91.1 0.24–0.41
Greater Pelvis 99.6 ± 0.7 0.16 ± 0.04 99.2 ± 1.1 0.22 ± 0.05 89.8 ± 6.5 0.44 ± 0.11
range 100.0–97.5 0.11–0.27 99.9–95.3 0.17–0.37 94.9–71.1 0.33–0.77