. 2000 Apr;38(4):1476–1481. doi: 10.1128/jcm.38.4.1476-1481.2000

TABLE 2.

Comparison of AxSYM CMV IgM assay results to consensus and resolved interpretationsa

Interpretation Serum sample group (no. of specimens) No. of specimens with the following interpretation:
% Sensitivity (95% CI) % Specificity (95% CI) % Agreement (95% CI)
Pos Eqv Neg None
Consensus Pos (93) 68 0 14 11 88.31 (78.97–94.51)
Eqv (21) 4 0 15 2 96.93 (94.90–98.31)
Neg (458) 9 1 442 6 95.68 (93.60–97.25)
Resolved Pos (93) 76 0 6 11 97.44 (91.04–99.69)
Eqv (21) 4 0 15 2 98.68 (97.16–99.52)
Neg (458) 2 1 450 5 98.50 (97.07–99.35)
a

Consensus assays: Abbott CMV-M EIA, Gull CMV IgM ELISA, and Trinity Biotech/Centocor CAPTIA CMV-M. Abbreviations: Pos, positive; Eqv, equivocal; Neg, negative; None, no interpretation. Sensitivity, specificity, and agreement are relative values for the consensus interpretation and resolved values for the resolved interpretation.