TABLE 1.

Correlation analysis and mean bias (%) of user performance, comparison with technician‐derived values, and comparison of sample types (capillary and venous blood samples) using the GCare Lipid Analyzer

Lipid test Correlation coefficient (R) Bias (%)
User performance TC 0.957 (0.938–0.971) −2.1
TG 0.991 (0.986–0.994) 1.5
HDL‐C 0.936 (0.908–0.955) 3.6
User vs. technician TC 0.963 (0.947–0.975) 2.0
TG 0.993 (0.989–0.995) 0.6
HDL‐C 0.939 (0.912–0.958) −2.4
Sample types TC 0.945 (0.928–0.957) −0.7
TG 0.988 (0.984–0.991) 0.0
HDL‐C 0.926 (0.904–0.942) −1.8

Abbreviations: HDL‐C, high‐density lipoprotein cholesterol; TC, total cholesterol; TG, triglyceride.