Skip to main content
. 2021 Nov 23;8:775299. doi: 10.3389/fmolb.2021.775299

TABLE 2.

Standard curve parameter in real-time digital PCR.

Target Slope Constant Rsqra Eff. (%)b Single-molecule region Transition region Bulk region
101 cp/pnl (occ.) c 102 cp/pnl (occ.) c 103 cp/pnl (occ.) c 104 cp/pnl (occ.) c 105 cp/pnl (occ.) c
blaIMP −2.953 37.875 0.978 118.111 7 (0.9%) 51 (6.6%) 519 (67.4%) 770 (100.0%) 768 (99.7%)
blaKPC −3.354 38.275 0.993 98.661 5 (0.6%) 56 (7.3%) 398 (51.7%) 769 (99.9%) 770 (100.0%)
blaNDM −3.705 40.62 0.996 86.174 4 (0.5%) 21 (2.7%) 190 (24.7%) 767 (99.6%) 769 (99.9%)
blaOXA-48 −3.304 38.01 0.998 100.77 3 (0.4%) 25 (3.2%) 321 (41.7%) 769 (99.9%) 768 (99.7%)
blaVIM −3.582 39.96 0.994 90.169 6 (0.8%) 59 (7.7%) 659 (85.6%) 770 (100.0%) 770 (100.0%)
a

R-squared.

b

Efficiency (%).

c

Copies/panel (% occupancy in digital PCR). The occupancy is calculated by counting the number of amplification reaction occurring per each panel and diving it by the total number of wells (N = 770).