Table 3.

Performance characteristics for real-time PCR duplexes.

Performance characteristic CT/hDNA duplex non-LGV/LGV duplex

Analytic sensitivity 93%/100% 100%/88%
Diagnostic sensitivity 93%/100% 100%/88%
Analytic specificity 100%/100% 100%/100%
Diagnostic specificity n/aa / n/aa 100%/100%
Precision/Reproducibility 100% 100%
Qualitative accuracy 93% 99%
Limit of detection (LOD) ≤100 copies/reaction ≤100 copies/reaction
Positive predictive value (PPV) 100% 100%
Negative predictive value (NPV) 84% 99%
a