. 2021 Jun 8;6(2):108–114. doi: 10.14744/bej.2021.28247

Table 3.

Summary of variance between the HandyRef-K-Retinomax, HandyRef-K-Plusoptix, Retinomax-Plusoptix, Table-top autorefractometer-Handyref-K, Table-top autorefractometer-Retinomax, and Table-top autorefractometer-Plusoptix measurements

Mean difference Standard deviation 95% limits of agreement P*
HandyRef-Retinomax
 SP -0.133 0.564 -0.2153-(-0.04992) 0.002
 CP 0.187 0.432 0.1231-(0.2499) p<0.0001
 Axis 0.690 9.082 -1.0026-(2.3831) 0.421
 SE -0.039 0.579 -0.1243-(0.04559) 0.362
HandyRef-Plusoptix
 SP 0.746 0.939 0.6082-(0.8835) p<0.0001
 CP 0.102 0.453 0.03573-(0.1687) 0.003
 Axis -0.115 11.583 -2.2739-(2.0439) 0.916
 SE 0.797 0.939 0.6593-(0.9346) p<0.0001
Retinomax- Plusoptix
 SP 0.878 0.986 0.7338-(1.0231) p<0.0001
 CP -0.084 0.282 -0.1256-(-0.04295) p= 0.0001
 Axis -0.805 11.543 -2.9569-(1.3463) p= 0.460
 SE 0.836 0.971 0.6940-(0.9787) p<0.0001
Table-top- HandyRef
 SP -0.30 0.56 -0.3833-(-0.2190) p<0.001
 CP -0.02 0.49D -0.09934-(0.04254) p= 0.431
 Axis 2 8.7 0.3814-(3.6186) p= 0.016
 SE -0.32 0.59 -0.4020-(-0.2287) p<0.001
Table-top-Retinomax
 SP -0.43 0.59 0.5196 –(-0.3479) p<0.001 *
 CP 0.16 0.30 0.1144-(0.2017) p<0.001
 Axis 2.69 9.6 0.930-(4.4776) p= 0.004
 SE -0.35 0.593 -0.4417-(-0.2677) p<0.001
Autorefractometer-PlusOptix A09
 SP 0.45 0.77 0.3314-(0.5580) p<0.001
 CP 0.07 0.36 0.02164-(0.1260) p= 0.006
 Axis 1.9 11.30 -0.2226-(3.9925) p= 0.079
 SE -0.48 0.74 0.3722-(0.5910) p<0.001
*

Bland-Altman analysis; CP: Cylindrical power; SE: Spherical equivalent; SP: Spherical power; HandyRef-K: Nidek Co. Ltd., Tokyo, Japan; Plusoptix A09: Plusoptix GmbH, Nuremberg, Germany; Retinomax K-plus 3: Right Mfg. Co. Ltd., Tokyo, Japan; Table-top autorefractometer: URK 800, Unicos Co. Ltd., Daejeon, Republic of Korea.