What is the future of electrical impedance spectroscopy in flow cytometry?

Furkan Gökçe 1,a), Paolo S Ravaynia 1, Mario M Modena 1, Andreas Hierlemann 1
PMCID: PMC8651262  PMID: 34917226

Abstract

More than 20 years ago, electrical impedance spectroscopy (EIS) was proposed as a potential characterization method for flow cytometry. As the setup is comparably simple and the method is label-free, EIS has attracted considerable interest from the research community as a potential alternative to standard optical methods, such as fluorescence-activated cell sorting (FACS). However, until today, FACS remains by and large the laboratory standard with highly developed capabilities and broad use in research and clinical settings. Nevertheless, can EIS still provide a complement or alternative to FACS in specific applications? In this Perspective, we will give an overview of the current state of the art of EIS in terms of technologies and capabilities. We will then describe recent advances in EIS-based flow cytometry, compare the performance to that of FACS methods, and discuss potential prospects of EIS in flow cytometry.

INTRODUCTION

In the early 2000s, electrical impedance spectroscopy (EIS) was introduced in flow cytometry for multi-parametric analysis of cells in suspension.1,2 The interest in EIS was fueled by its potential to become a low-complexity and low-cost alternative to established fluorescence- and scattering-based detection methods, which were commonly employed in flow cytometry, the so-called fluorescence-activated cell sorting (FACS) [Fig. 1(a)]. Inspired by the Coulter counter principle,3 EIS-based detection methods for flow cytometry (EIS-FC) measure variations in electrical impedance, while a flow of medium including cells passes over or in between a set of electrodes in contact with the suspension medium.2,4,5 EIS-FC uses a microfluidic channel to align the cells for single-cell detection and to focus them within the sensing volume, an approach very similar to that employed for focusing cells in the detection volume of a FACS cytometer [Fig. 1(b)]. The microfluidic channel provides a small sample volume between the electrodes (∼10 nl) for increased sensitivity, and the laminar flow profile enables us to precisely control the cell flow.1,6 Similar information on the cells in suspension can be obtained by using either label-free FACS-based or EIS-based flow cytometry. Label-free FACS analysis relies on the detection of forward- and side-scattered light (FSC-A and SSC-A) to get information on cell size and cell granularity [Fig. 1(a)]. By recording cell impedance at multiple frequencies to comprehensively assess cell dielectric properties, EIS-FC provides estimates of the cell size at low frequencies (usually around or below 100 kHz) or information on membrane integrity and intracellular content by looking at the impedance at high frequencies (from ∼100 kHz to ∼10–100 MHz) [Fig. 1(c)].2 High-frequency impedance recordings are then often normalized with respect to low-frequency impedance values to be able to compare information on the membrane integrity of cells that feature different geometric dimensions or sizes. The ratio of high-frequency impedance to low-frequency impedance values is referred to as “opacity” [Fig. 1(b)].2,7,8

FIG. 1.

FIG. 1.

Working principles of (a) FACS-based flow cytometry and (b) EIS-based flow cytometry, showing the application and recording of electrical AC signals. In the subpanels are typical FACS and EIS scatterplots of a hypothetical mixture of two populations, which show similarities for label-free analysis. In a typical FACS scatterplot, debris is visible besides the cell populations, while it is generally not plotted in EIS scatterplots. Opacity, as shown along the y axis of the EIS scatterplot, denotes the ratio of the high- and low-frequency signal amplitude. It is used to minimize dependence of EIS signals on the relative positions of the measured cells. Opacity can be used to characterize membrane integrity and intracellular contents of the cells. (c) Top: equivalent-circuit model of the sensing region, showing the contributions of the electrodes, of the solution, and of the passing cell to the sensed impedance. Constant phase elements (CPE) represent the electrodes. Bottom: signal-magnitude and phase-response plots of cells in dependence of the applied frequency.

When EIS-FC was initially developed, FACS instruments required frequent re-alignment of the optical elements and highly trained personnel for data acquisition and analysis. In contrast, the sensing electrodes of EIS-FC could be integrated within the microfluidic channel, so that no user alignment was necessary. Furthermore, EIS signal detection could be carried out with small and portable electronic equipment, which resulted in a small footprint of the detection system and low maintenance and acquisition costs. Therefore, EIS-FC initially raised considerable interest in the scientific community, and many implementations have been developed with the aims to increase sensitivity, simplify electrode integration, and extend the information that can be obtained through EIS by using refined electrical-equivalent models of cells in an electrolyte flow.5,9–11

IS EIS-BASED FLOW CYTOMETRY A VIABLE ALTERNATIVE TO FACS?

EIS-FC has proven to be competitive with FACS-based cytometry in multiple aspects. By using state-of-the-art electronic components and by parallelizing the analysis, EIS-FC can readily be used to record thousands of events per second,10,12,13 which matches the throughput of FACS-based analysis.14 FACS detection has been optimized for the analysis of single mammalian cells in the 1–100 μm diameter range,14 while ad hoc modifications of the instrumentation can be made to extend the measurement range from sub-micrometer15–17 to millimeter-range particles.18,19 However, such modifications require significant efforts and are costly. In contrast, EIS detection can be easily adapted for a wide range of dimensions of analyzed particles—from sub-micrometer particles (e.g., extracellular vesicles or apoptotic bodies) to large cell aggregates or organisms (e.g., organoids or parasites)—with minimal modifications to the measurement equipment. Only comparably simple modifications of the detection chamber and a careful selection of the frequency range of the impedance signals that will be recorded are required.20–27

The main limitation of EIS-FC originates from its limited capability in identifying different cell types in a mixed population as a consequence of the label-free nature of cell-size and membrane-property measurements. In contrast, great specificity is provided by fluorescent labels of different cellular markers that enable us to characterize mixed cell populations at high resolution by using FACS. FACS techniques have experienced drastic improvements in instrumentation and in the development of novel labeling strategies for fluorescence-based cell characterization,28 which have greatly improved its performance and user-friendliness. New-generation FACS instruments require minimal or no alignment of the optical elements and can be readily used for automated high-throughput detection. Furthermore, a comprehensive analysis can now be carried out with only a few microliters of the sample. Modern FACS devices that include multiple laser sources can simultaneously detect up to 20 parameters for characterizing and differentiating cells in mixed populations. The use of quantum-dot-based cellular labels could enable FACS detection by using up to ∼50 different markers.14 The analysis of these multidimensional FACS data has also been greatly facilitated by the development of intuitive and automated analysis software packages. The integration of machine-learning approaches for automated clustering is expected to further improve the analysis performance and simplify data interpretation.14,29 Improvements in FACS technology have promoted its adoption in research, industrial, and clinical settings despite the high initial costs to purchase the equipment.12,14,30,31 A disruptive recent advancement in flow cytometry includes the integration of high-speed cameras for high-resolution imaging in FACS flow cytometers.32–34 This new method, which has been termed image-activated cell sorting (IACS) or imaging flow cytometry (IFC), provides spatial information on the distribution of labeled markers in single cells in addition to the characterization by FACS-based analysis. IACS provides more information and higher data dimensionality and enables in-depth characterization and refined differentiation of cells in mixed populations. An overview of the most important features of EIS-FC, FACS, and IACS is given in Table I.

TABLE I.

Summary of the main features of EIS-FC, FACS, and IACS.

EIS-FC FACS IACS
Installation costs <$100k $50k–$1M >$1M
Throughput in events per second Tens of thousands of cells per second Tens of thousands of cells per second Around 100 cells per second
Dimensionality of data Defined by the number of applied and recorded electrical signal frequencies Around 20 features for a standard FACS device High-dimensional data due to spatial information provided through imaging
Integration
and multiplexing
Convenient integration and scalability due to electronic multiplexing Limited and expensive due to setup costs and optical alignment Limited due to complex optics, setup costs, and optical alignment
Label-free Yes Yes + fluorescence Yes + fluorescence
Real-time or end point Both Limited to end point Limited to end point

Although the application of EIS-FC requires significantly less investments in terms of setup costs and efforts for operation, this electrical analysis method has been commercially developed to a much lesser extent than optical methods and has been applied mostly to cell counting.35 Meanwhile, FACS devices and FACS analysis have been developed at a fast pace and are broadly available to researchers and clinicians. Therefore, an obvious question concerns the potential future of EIS-based detection methods in flow cytometry.

WHAT IS THE FUTURE OF EIS-BASED FLOW CYTOMETRY?

Flow cytometry is predominantly used as an end point analysis method to characterize and analyze cell populations, where the performance of FACS analysis is superior to that of EIS-FC. The single-cell resolution of flow cytometry methods could, however, also provide a wealth of information in cell cultures and could be used to monitor dynamic changes in real time. The large form factor of FACS equipment prevents integration in compact systems and cell-culture platforms. In contrast, EIS methods could be an ideal candidate for continuous characterization of a wide range of biological samples, as they are conveniently configurable, label-free, and can easily be automated and multiplexed.24,26 Application examples include monitoring responses upon drug exposure. The parallel and continuous EIS monitoring of cancer microtissues upon exposure to drug compounds was recently demonstrated.21 The microtissues were periodically moved across EIS electrodes by simple tilting of the culture platform, which initiated medium flow and enabled parallel impedance screening of multiple microtissues in real time. EIS-FC could potentially be used in a similar way for label-free real-time monitoring of suspended cells at single-cell resolution.

Commercial impedance-based Coulter counter systems have seen great technological improvements in the last few years and feature high throughput and a large detection range spanning from sub-micrometer particle diameters to a few millimeters. They have been used for the characterization of a broad variety of samples including body fluids.36–39 The label-free electrical detection principle obviates long sample preparation times and the need for expensive dyes, which renders EIS systems particularly interesting for diagnostic and medical purposes. Academic research efforts have yielded micro-fabricated devices with distinct advantages over commercial impedance-based cytometers, which include reduced sample volumes, low power consumption, and small system dimensions and portability.40

The compact nature and high level of integration of EIS-FCs also enable application as in-line analysis techniques for quality control or monitoring of cell-manipulation protocols in systems with integrated electrodes. For instance, EIS-FC could be used to monitor the effectiveness of cell electroporation in flow-through electroporators or for optimization of electroporation parameters.41,42 Similarly, integrated electrodes for EIS-FC could be utilized in droplet microfluidics for sorting and analyzing encapsulated cells43–45 or for manipulating droplets.46,47 Finally, EIS-FC could also find applications in quality assessment of bioprinting of cells and organoids to verify, in an automated and high-throughput manner, size uniformity and viability.48–50

CONCLUSION

In conclusion, although EIS can still be used as an end point analysis method in flow cytometry as suggested two decades ago, the technological advances in FACS and the recent introduction of optical microscopy information in IACS flow cytometry render the optical techniques superior for cell population measurements. However, the label-free nature and high flexibility in configuring and integrating EIS methods may open a variety of applications for EIS methods. Examples include the real-time monitoring of compound effects in long-term drug exposure experiments or the in-line monitoring of cells and organoids for quality control purposes during bioprinting. Finally, the possibility of realizing small-outline and low-cost instrumentation renders EIS-based methods to be very attractive in resource-limited settings or for analyzing body fluids in the growing field of wearable medical devices.

ACKNOWLEDGMENTS

The authors thank Dr. Aleksandra Gumienny for her valuable input on FACS. This work was financially supported by the Swiss National Science Foundation under Contract No. CR32I2_166329: “Infected body-on-chip.” F.G. acknowledges individual support through the Personalized Health and Related Technologies (PHRT) initiative of the ETH Domain (Project No. SFA-PHRT-2017-309).

AUTHOR DECLARATIONS

Conflict of Interest

The authors have no conflicts to disclose.

DATA AVAILABILITY

Data sharing is not applicable to this article as no new data were created or analyzed in this study.

REFERENCES

  • 1.Ayliffe H. E., Frazier A. B., and Rabbitt R. D., “Electric impedance spectroscopy using microchannels with integrated metal electrodes,” J. Microelectromech. Syst. 8, 50–57 (1999). 10.1109/84.749402 [DOI] [Google Scholar]
  • 2.Gawad S., Schild L., and Renaud Ph., “Micromachined impedance spectroscopy flow cytometer for cell analysis and particle sizing,” Lab Chip 1, 76 (2001). 10.1039/b103933b [DOI] [PubMed] [Google Scholar]
  • 3.Coulter W. H., “Means for counting particles suspended in a fluid,” U.S. patent 2,656,508 (20 Oct 1953).
  • 4.Gawad S., Cheung K., Seger U., Bertsch A., and Renaud P., “Dielectric spectroscopy in a micromachined flow cytometer: Theoretical and practical considerations,” Lab Chip 4, 241 (2004). 10.1039/b313761a [DOI] [PubMed] [Google Scholar]
  • 5.Sun T. and Morgan H., “Single-cell microfluidic impedance cytometry: A review,” Microfluid. Nanofluid. 8, 423–443 (2010). 10.1007/s10404-010-0580-9 [DOI] [Google Scholar]
  • 6.Cheung K. C. et al. , “Microfluidic impedance-based flow cytometry,” Cytometry Part A 77A, 648–666 (2010). 10.1002/cyto.a.20910 [DOI] [PubMed] [Google Scholar]
  • 7.Coulter W. H. and Hogg W. R. A., “Collating apparatus for pairs of electrical pulses produced by particle analyzing apparatus,” U.S. patent 3,502,973 (24 March 1970).
  • 8.Hoffman R. A., Johnson T. S., and Britt W. B., “Flow cytometric electronic direct current volume and radiofrequency impedance measurements of single cells and particles,” Cytometry 1, 377–384 (1981). 10.1002/cyto.990010605 [DOI] [PubMed] [Google Scholar]
  • 9.Zhu S. et al. , “Microfluidic impedance cytometry for single-cell sensing: Review on electrode configurations,” Talanta 233, 122571 (2021). 10.1016/j.talanta.2021.122571 [DOI] [PubMed] [Google Scholar]
  • 10.Honrado C., Bisegna P., Swami N. S., and Caselli F., “Single-cell microfluidic impedance cytometry: From raw signals to cell phenotypes using data analytics,” Lab Chip 21, 22–54 (2021). 10.1039/D0LC00840K [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Petchakup C. et al. , “Advances in single cell impedance cytometry for biomedical applications,” Micromachines 8, 87 (2017). 10.3390/mi8030087 [DOI] [Google Scholar]
  • 12.Yang R.-J., Fu L.-M., and Hou H.-H., “Review and perspectives on microfluidic flow cytometers,” Sens. Actuators B Chem. 266, 26–45 (2018). 10.1016/j.snb.2018.03.091 [DOI] [Google Scholar]
  • 13.Liao Z. et al. , “Recent advances in microfluidic chip integrated electronic biosensors for multiplexed detection,” Biosens. Bioelectron. 121, 272–280 (2018). 10.1016/j.bios.2018.08.061 [DOI] [PubMed] [Google Scholar]
  • 14.Manohar S. M., Shah P., and Nair A., “Flow cytometry: Principles, applications and recent advances,” Bioanalysis 13, 185–198 (2021). 10.4155/bio-2020-0267 [DOI] [PubMed] [Google Scholar]
  • 15.Brussaard C. P. D., Marie D., and Bratbak G., “Flow cytometric detection of viruses,” J. Virol. Methods 85, 175–182 (2000). 10.1016/S0166-0934(99)00167-6 [DOI] [PubMed] [Google Scholar]
  • 16.Steen H. B., “Flow cytometer for measurement of the light scattering of viral and other submicroscopic particles,” Cytometry 57A, 94–99 (2004). 10.1002/cyto.a.10115 [DOI] [PubMed] [Google Scholar]
  • 17.Arkesteijn G. J. A., Lozano-Andrés E., Libregts S. F. W. M., and Wauben M. H. M., “Improved flow cytometric light scatter detection of submicron-sized particles by reduction of optical background signals,” Cytometry Part A 97, 610–619 (2020). 10.1002/cyto.a.24036 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Watson D. A., Gaskill D. F., Brown L. O., Doorn S. K., and Nolan J. P., “Spectral measurements of large particles by flow cytometry,” Cytometry Part A 75A, 460–464 (2009). 10.1002/cyto.a.20706 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Randall C. J. et al. , “Rapid counting and spectral sorting of live coral larvae using large-particle flow cytometry,” Sci. Rep. 10, 12919 (2020). 10.1038/s41598-020-69491-0 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Haandbæk N., Bürgel S. C., Heer F., and Hierlemann A., “Characterization of subcellular morphology of single yeast cells using high frequency microfluidic impedance cytometer,” Lab Chip 14, 369–377 (2014). 10.1039/C3LC50866H [DOI] [PubMed] [Google Scholar]
  • 21.Bürgel S. C., Diener L., Frey O., Kim J.-Y., and Hierlemann A., “Automated, multiplexed electrical impedance spectroscopy platform for continuous monitoring of microtissue spheroids,” Anal. Chem. 88, 10876–10883 (2016). 10.1021/acs.analchem.6b01410 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Haandbæk N., With O., Bürgel S. C., Heer F., and Hierlemann A., “Resonance-enhanced microfluidic impedance cytometer for detection of single bacteria,” Lab Chip 14, 3313–3324 (2014). 10.1039/C4LC00576G [DOI] [PubMed] [Google Scholar]
  • 23.Honrado C. et al. , “Apoptotic bodies in the pancreatic tumor cell culture media enable label-free drug sensitivity assessment by impedance cytometry,” Adv. Biol. 5, 2100438 (2021). 10.1002/adbi.202100438 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Spencer D. C. et al. , “A fast impedance-based antimicrobial susceptibility test,” Nat. Commun. 11, 5328 (2020). 10.1038/s41467-020-18902-x [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Ravaynia P. S. et al. , “Parallelized impedance-based platform for continuous dose-response characterization of antischistosomal drugs,” Adv. Biosyst. 4, 1900304 (2020). 10.1002/adbi.201900304 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Chawla K. et al. , “Integrating impedance-based growth-rate monitoring into a microfluidic cell culture platform for live-cell microscopy,” Microsyst. Nanoeng. 4, 1–12 (2018). 10.1038/s41378-018-0006-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Tang W., Tang D., Ni Z., Xiang N., and Yi H., “Microfluidic impedance cytometer with inertial focusing and liquid electrodes for high-throughput cell counting and discrimination,” Anal. Chem. 89, 3154–3161 (2017). 10.1021/acs.analchem.6b04959 [DOI] [PubMed] [Google Scholar]
  • 28.McKinnon K. M., “Flow cytometry: An overview,” Curr. Protoc. Immunol. 120, 5.1.1 (2018). 10.1002/cpim.40 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Gong Y., Fan N., Yang X., Peng B., and Jiang H., “New advances in microfluidic flow cytometry,” Electrophoresis 40, 1212–1229 (2019). 10.1002/elps.201800298 [DOI] [PubMed] [Google Scholar]
  • 30.Jaye D. L., Bray R. A., Gebel H. M., Harris W. A. C., and Waller E. K., “Translational applications of flow cytometry in clinical practice,” J. Immunol. 188, 4715–4719 (2012). 10.4049/jimmunol.1290017 [DOI] [PubMed] [Google Scholar]
  • 31.Doan M. et al. , “Diagnostic potential of imaging flow cytometry,” Trends Biotechnol. 36, 649–652 (2018). 10.1016/j.tibtech.2017.12.008 [DOI] [PubMed] [Google Scholar]
  • 32.Nitta N. et al. , “Intelligent image-activated cell sorting,” Cell 175, 266–276.e13 (2018). 10.1016/j.cell.2018.08.028 [DOI] [PubMed] [Google Scholar]
  • 33.Isozaki A. et al. , “A practical guide to intelligent image-activated cell sorting,” Nat. Protoc. 14, 2370–2415 (2019). 10.1038/s41596-019-0183-1 [DOI] [PubMed] [Google Scholar]
  • 34.Weiss L. E. et al. , “Three-dimensional localization microscopy in live flowing cells,” Nat. Nanotechnol. 15, 500–506 (2020). 10.1038/s41565-020-0662-0 [DOI] [PubMed] [Google Scholar]
  • 35.Vembadi A., Menachery A., and Qasaimeh M. A., “Cell cytometry: Review and perspective on biotechnological advances,” Front. Bioeng. Biotechnol. 7, 147 (2019). 10.3389/fbioe.2019.00147 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Strohm E. M. et al. , “Sizing biological cells using a microfluidic acoustic flow cytometer,” Sci. Rep. 9, 4775 (2019). 10.1038/s41598-019-40895-x [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Martinelli C. et al. , “Development of artificial plasma membranes derived nanovesicles suitable for drugs encapsulation,” Cells 9, 1626 (2020). 10.3390/cells9071626 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Martín M. J. A. et al. , “Automated cell count in body fluids: A review,” Adv. Lab. Med. Av. Med. Lab. 2, 149–161 (2021). 10.1515/almed-2021-0011 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Heidmann I., Schade-Kampmann G., Lambalk J., Ottiger M., Berardino M. D., and Wang T., “Impedance flow cytometry: A novel technique in pollen analysis,” PLoS One 11, e0165531 (2016). 10.1371/journal.pone.0165531 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Furniturewalla A., Chan M., Sui J., Ahuja K., and Javanmard M., “Fully integrated wearable impedance cytometry platform on flexible circuit board with online smartphone readout,” Microsyst. Nanoeng. 4, 1–10 (2018). 10.1038/s41378-018-0019-0 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Ye Y. et al. , “Single-cell electroporation with real-time impedance assessment using a constriction microchannel,” Micromachines 11, 856 (2020). 10.3390/mi11090856 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Buchmann L., Frey W., Gusbeth C., Ravaynia P. S., and Mathys A., “Effect of nanosecond pulsed electric field treatment on cell proliferation of microalgae,” Bioresour. Technol. 271, 402–408 (2019). 10.1016/j.biortech.2018.09.124 [DOI] [PubMed] [Google Scholar]
  • 43.Yakdi N. E., Huet F., and Ngo K., “Detection and sizing of single droplets flowing in a lab-on-a-chip device by measuring impedance fluctuations,” Sens. Actuators B Chem. 236, 794–804 (2016). 10.1016/j.snb.2016.05.123 [DOI] [Google Scholar]
  • 44.Panwar J. and Roy R., “Integrated field’s metal microelectrodes based microfluidic impedance cytometry for cell-in-droplet quantification,” Microelectron. Eng. 215, 111010 (2019). 10.1016/j.mee.2019.111010 [DOI] [Google Scholar]
  • 45.Fan W. et al. , “Single-cell impedance analysis of osteogenic differentiation by droplet-based microfluidics,” Biosens. Bioelectron. 145, 111730 (2019). 10.1016/j.bios.2019.111730 [DOI] [PubMed] [Google Scholar]
  • 46.Teo A. J. T. et al. , “Controllable droplet generation at a microfluidic T-junction using AC electric field,” Microfluid. Nanofluid. 24, 1–9 (2020). 10.1007/s10404-019-2306-y [DOI] [Google Scholar]
  • 47.Doonan S. R. and Bailey R. C., “K-channel: A multifunctional architecture for dynamically reconfigurable sample processing in droplet microfluidics,” Anal. Chem. 89, 4091–4099 (2017). 10.1021/acs.analchem.6b05041 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48.Schoendube J., Wright D., Zengerle R., and Koltay P., “Single-cell printing based on impedance detection,” Biomicrofluidics 9, 014117 (2015). 10.1063/1.4907896 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Gong L. et al. , “Direct and label-free cell status monitoring of spheroids and microcarriers using microfluidic impedance cytometry,” Small 17, 2007500 (2021). 10.1002/smll.202007500 [DOI] [PubMed] [Google Scholar]
  • 50.Hong G. et al. , “Production of multiple cell-laden microtissue spheroids with a biomimetic hepatic-lobule-like structure,” Adv. Mater. 33, 2102624 (2021). 10.1002/adma.202102624 [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

Data sharing is not applicable to this article as no new data were created or analyzed in this study.