Skip to main content
. 2021 Aug 24;17(8):1010–1017. doi: 10.2174/1573405617666210126155814

Table 3.

SNR and CNR values for TSE imaging of lumbar spine without and with CS.

- T1WI T2WI PDWI
CS SENSE P-values CS SENSE P-values CS SENSE P-values
SNR(bone marrow) 15.50±4.30 15.58±4.18 0.977 16.17±3.20 16.80±8.83 0.885 7.33±1.58 10.21±3.48 0.131
SNR(muscle) 29.40±8.65 29.13±8.51 0.962 12.07±1.51 11.63±0.99 0.598 13.22±2.66 14.65±3.11 0.456
SNR(CSF) 10.55±3.13 13.14±2.54 0.188 72.20±8.51 67.41±7.55 0.374 51.11±4.80 52.30±5.74 0.731
SNR(spinal cord) 17.88±5.36 21.64±5.27 0.295 24.91±0.82 25.62±3.16 0.639 32.56±8.55 38.53±7.94 0.286
SNR(nerve root) 7.62±1.78 11.30±4.95 0.157 - - - - - -
CNR(bone marrow /CSF) 9.36±2.93 9.58±2.35 0.900 27.97±7.60 25.80±8.66 0.686 37.05±7.68 34.01±5.38 0.489
CNR(muscle /CSF) 6.80±1.04 7.75±1.70 0.319 52.15±5.45 48.20±1.47 0.156 44.23±3.73 43.32±4.45 0.736
CNR(spinal cord /CSF) 6.03±2.22 10.02±3.86 0.080 37.15±3.18 35.12±2.75 0.312 31.68±4.31 29.79±2.46 0.419
CNR(nerve root /CSF) 2.63±0.97 4.06±2.07 0.201 - - - - - -

SNR = signal-to-noise ratio, CNR = contrast-to-noise ratio, CSF = cerebrospinal fluid.

The data are presented as means ± standard deviations.