TABLE 2.

Comparison table of MIC values originating of microdilution and MICya

Broth Bacteria Measurement Essential agreement Discrepancy resolution testing EA Weighted Fleiss’ kappa ± SE
Both All 432 340 (78.7%) 83.3% 0.714
LB Gram positive 108 88 (81.5%) 69.4% 0.66
Gram negative 108 79 (73.1%) 86.1% 0.731
LB sum Both 216 167 (77.3%) 77.8% 0.706
MH Gram positive 108 83 (76.9%) 83.3% 0.662
Gram negative 108 90 (83.3%) 94.4% 0.76
MH sum Both 216 173 (80.1%) 88.9% 0.72
a

Essential agreement represents the number and percent of MIC values originating from MICy that were in ±1 2-fold dilution range to the reference method. Essential agreement percent of discrepancy resolution testing was calculated according to ISO/DIS 20776-2:2021 standard.