1 |
The device has a stated accuracy of ± 0.3°C or lower |
◽ |
2 |
The device has a minimum resolvable temperature difference (sensitivity) of 0.1°C. This is often stated as the “resolution” of spot infrared thermometers |
◽ |
|
Purchasing a thermal imaging camera |
|
3 |
The total resolution of the raw image is at-least 320 × 240 image pixels |
◽ |
4 |
The device makes drift compensation (self-corrections) to keep the camera within the specified accuracy |
◽ |
|
If no to above |
|
5 |
An external temperature reference source (ETRS, black body calibrator) is used with the device. The ETRS will have a combined stability and drift error of 0.1°C or less over a 14-day period. The ETRS will have an expanded uncertainty below ± 0.3°C |
◽ |
|
Measurement location |
|
6 |
Air temperature is stable, and between 18 and 24°C |
◽ |
7 |
Air relative humidity is stable, and between 20 and 75% |
◽ |
8 |
The room is free of forced air movement and sources of radiation (i.e., no reflective backgrounds, sunlight) that can impact the NCIT display temperature. A non-reflective cloth backdrop can be used to minimize sources of radiation. |
◽ |
9 |
The room is at-least 2 × 3 meters in size |
◽ |
|
Device operation |
|
10 |
The user has consulted the manual to determine the correct measuring location and distance from the subject. |
◽ |
11 |
The device is not overdue a calibration (see manufacturer guidelines) |
◽ |
12 |
The device emissivity is set to 0.98 for long wave IR cameras or devices |
◽ |
13 |
The device has warmed up to room temperature for at least 30 minutes |
◽ |
14 |
The device is taking a measurement from a 90° angle and perpendicular to the front of the subjects’ face |
◽ |
15 |
For spot measurements, the target is specified as the center of the forehead |
◽ |
16 |
For thermal imaging, the target is specified as the inner eye-canthus (i.e., inner corner of the eyelid) |
◽ |
17 |
For thermal imaging, the rainbow color scale is used |
◽ |
18 |
An absolute temperature threshold has been established through consultation with a healthcare professional or expert in the field of fever screening with NCIT |
◽ |
19 |
Masks, eyeglasses, hats, and any other items obstructing the face have been removed |
◽ |
|
If using an ETRS |
|
20 |
The ETRS in focus and large enough to be easily discriminated from the background |
◽ |
21 |
The ETRS is no larger than 10% of the size of subject’s face (in a face only image) |
◽ |
22 |
The ETRS is set to a temperature close to the threshold temperature (i.e., 35–38°C) |
◽ |
|
Reporting |
|
23 |
Model/manufacturer of camera; accuracy, resolution, model of ETRS, lens type, software & emissivity used |
◽ |
24 |
Location, environment (air temperature & humidity), Reflected temperature |
◽ |
25 |
Any pre- ‘treatment of subjects scanned (i.e., acclimation to conditions) |
◽ |