Abstract
Originally, the future of automated insulin delivery (AID) systems, or artificial pancreas systems (APS), was having them at all, in any form. We’ve learned in the last half dozen years that the future of all artificial pancreas systems holds higher time in range, less work required to manage automated insulin delivery systems to improve quality of life, and the ability to input critical information back into the system itself. The data and user experience stories make it clear: APS works. APS are an improvement over other diabetes therapy methods when they are made available, accessible, and affordable. Understanding the unmet expectations of current users of first generation APS technology may also aid in the development of improved technology and user experiences for the future of APS.
Keywords: artificial pancreas, APS, closed loop, automated insulin delivery, AID, hybrid closed loop
Introduction
In the early 2010s, the future of artificial pancreas systems or automated insulin delivery systems was simply having them at all. I know this well, as I was one of the first people to build my own hybrid closed loop system in December 2014. 1 In subsequent years, the first hybrid closed loop commercial systems have become available in some countries around the world, although more work is needed to be done to make them universally available 2 (and affordable and accessible). And, over those same years, the system I first used and made open source became used by thousands of individuals 3 around the world. What we’ve learned in the open source community among those who have chosen to “do it yourself” (or “DIY”) is what the future of all artificial pancreas systems holds: higher time in range, less work required to manage automated insulin delivery systems to improve quality of life, and the ability to input critical information back into the system itself.
“How it Started” for Artificial Pancreas Systems
The early results for all artificial pancreas systems, also known as hybrid closed loops or automated insulin delivery systems, is fairly similar. Time in range (TIR) typically starts low for manual diabetes treatment methods, even in “highly motivated” populations, and increases once people use this technology. This is true for the earliest studies of open source systems, where the first self-reported study showed TIR increase from 58 to 81% 4 and later analyses of larger populations also showed similar increased TIR.5-8 These open source closed loop systems reduce hyperglycemia without increasing hypoglycemia 9 and improve quality of life. 10 They have been studied not only in retrospective and observational studies11,12 but also in silico 13 as well as ongoing, randomized control trials. 14
Similarly, commercial closed loop systems 15 have shown the ability to increase time in range for people living with diabetes. Compared with manual control (where the person with diabetes is doing all decision making), commercial automated insulin delivery systems also typically increase time in range. In some cases, TIR increases upwards of 21 percentage points. 16 More recent systematic reviews and meta-analyses with data as recent as 2018 also show significant improvement up to 9.6 percentage points. 17
The data and user experience stories make it clear: artificial pancreas systems work, when they are made available, accessible, and affordable, and are an improvement over other diabetes therapy methods. 18 However, expectations 19 of the individual using the automated insulin delivery system are key. Understanding the unmet expectations of current users20,21 of current APS technology may also aid in the development of improved technology and user experiences for the future of APS.
“How it’s Going” for the Most Advanced Artificial Pancreas Systems
It’s possible to get more than 10 percentage point improvements in TIR out of insulin-only artificial pancreas systems. And it’s possible to achieve these TIR gains with minimal burden on the end user of the system.
For example, the open source community initially focused on using temporary basal adjustments for automating insulin delivery. This was done for safety reasons, 22 so that if the insulin pump lost connectivity with the system, the temporary basal rate would expire and the pump would resume standalone pump therapy. Later, however, it was determined that it would be possible to add boluses to the algorithm’s ability to adjust insulin delivery. The method was named “supermicroboluses” (known as SMB), 23 because it was a miniature version of the “superbolus” approach to “front-shift” insulin activity. It was not a synonym for a no-bolus, fully closed loop automated insulin delivery system; however, combined with other features such as the “unannounced meal” (UAM) feature, it became possible for some individuals to achieve a fully closed loop, that did not require user input for meals, while achieving the same TIR as before.
Some makers of commercial systems are working on a full closed loop system as well. 24 Some are experimenting with “no carbohydrate counting”—which does not mean zero meal announcement, but may include a meal announcement that requires roughly categorizing meal size. Others are experimenting with no announcement at all. Depending on the speed of insulin that is used in the system, as well as the algorithm, users may get different results in terms of meal outcomes and TIR with a “fully closed loop system.” With future faster insulins, it will be easier for a full closed loop system to respond to unexpected increases in blood glucose levels, regardless of the cause, and with a shorter insulin tail, further minimize the risk of hypoglycemia as well.
In the end, it comes down to a user and what they prefer, and allowing them to choose between their tradeoffs:
No need to bolus
No need to count carbs and/or announce meals
The ability to consume medium to higher carbohydrate meals
80%+ time in range
No hypoglycemia.
We believe, and have demonstrated in the open source community, that it’s possible to choose any combination of 4 of the 5—and with the latest, fastest insulins, even all 5 are possible. 25
This is what we should be striving for with the future of artificial pancreas systems. Settling for only 1 or 2 of these items (such as 70%-80% TIR and no or limited hypoglycemia) is the current status quo.
What’s in a Name? Microbolus, Correction Bolus, and More—It Doesn’t Matter
It ultimately doesn’t matter what you call various components and features of the closed loop systems, but it does matter that we work to make it clear to end users—people living with diabetes—what the capabilities of the systems are, and what they can or cannot do.
For example, much attention is given to whether a “correction bolus” can be issued by a closed loop system. That is perceived to be different than “microbolus(es).”
What matters, ultimately, though, is how much insulin is allowed to be delivered by the system; when and how often; and what the target is. That information determines how “aggressive” a system can be for correcting or preventing hyperglycemia.
What’s Going Away: A Perception of a Fixed “Correct” Basal to Bolus Ratio
Similarly, an artifact of past diabetes treatment methodologies has been a perception of the “correct” ratio of basal to bolus insulin, as a proxy of determining whether someone’s settings in their pump are “correct.” In an automated insulin delivery system, where the algorithm determines whether to issue a “basal” adjustment or a “bolus correction” (or microbolus, or correction bolus, etc.), the “choices” of type of insulin delivery is no longer a reflection on the user. Additionally, there are new tools such as Autotune 26 that can be used (and should be further studied and adapted by commercial companies, to better improve use of their closed loop systems) to adjust a person’s baseline basal rates, ISF, and carb ratio, whether that’s used by the closed loop system or whether that information is only used when a user “falls back” to manual mode.
What Needs to be Added to The Future of APS: Flexible User Inputs
One of the benefits—and detriments—of open source automated insulin delivery systems is the plethora of customization and input options. On the other end of the scale, with very few user input options beyond carbohydrate counting, are current first-generation commercial closed loop systems. The future of APS is somewhere in between these two.
While a typical person with diabetes doesn’t need all of the customization options available in DIY, there is pent up frustration in the diabetes community with the first generation of commercial APS technology. Because of the timing of insulin, and a person’s knowledge of their own behavior, it is important to be able to interact with a closed loop system and give it information that it can’t possibly predict, such as whether someone is about to eat or exercise. But beyond carbohydrate entries and slight target adjustments, the future of APS needs to involve and develop a few more, carefully designed user inputs.
For example, a person with diabetes needs the ability to tell the system that it has learned something “wrong,” such as when it detects a pattern that no longer exists. 27 Similarly, a person may have taken action to adjust for thing that the system has learned. For example, if the system is increasing insulin because it perceives the user to be more resistant, and the reason for resistance is resolved by a pump site change, the user needs to be able to inform the system to revert back to neutral or to otherwise stop applying whatever the system has “learned.”
These user inputs don’t need to be things that users input all the time, but they need to be available for the once a week or once a month occurrences that can otherwise often derail diabetes management, and leave users frustrated if the system can’t adapt quickly enough because it doesn’t know about the situation, or because it’s too slow to learn the situation has reverted back to normal.
Conclusion
At the end of the day, the future of artificial pancreas systems is already here: it is just not distributed widely enough yet. We know it is possible to achieve 80%-90%+ time in range without regular meal announcement or precise carbohydrate counting by end users, while still preventing hypoglycemia, reducing burden and improving quality of life for individuals living with diabetes who want to use these systems.
Footnotes
Abbreviations: AID, automated insulin delivery; APS, artificial pancreas/artificial pancreas system; CGM, continuous glucose monitor; DIY, do-it-yourself; TIR, time in range.
Declaration of Conflicting Interests: The author declared the following potential conflicts of interest with respect to the research, authorship, and/or publication of this article: While no financial conflict of interest exists, the researcher acknowledges she is one of the creators of and contributors to one of the open-source automated insulin delivery systems (OpenAPS).
Funding: The author received no financial support for the research, authorship, and/or publication of this article.
ORCID iD: Dana Lewis
https://orcid.org/0000-0001-9176-6308
References
- 1. Lewis D. History and perspective on DIY closed looping. J Diabetes Sci Technol. 2019;13:793-793. doi: 10.1177/1932296818808307 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Automated insulin delivery. Accessed May 14, 2021. https://diatribe.org/automated-insulin-delivery
- 3. OpenAPS Outcomes. Accessed March 30, 2021. https://openaps.org/outcomes/
- 4. Lewis D, Leibrand S. Real-world use of open source artificial pancreas systems. J Diabetes Sci Technol. 2016;10(6):1411. doi: 10.1177/1932296816665635 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Melmer A, Züger T, Lewis D, Leibrand S, Stettler C, Laimer M. Glycemic control in individuals with type 1 diabetes using an open source artificial pancreas system (OpenAPS). Diabetes Obes Metab. 2019;21:2333-2337. doi: 10.1111/dom.13810 [DOI] [PubMed] [Google Scholar]
- 6. Asarani NAM, Reynolds AN, Elbalshy M, et al. Efficacy, safety, and user experience of DIY or open-source artificial pancreas systems: a systematic review. Acta Diabetol. 2021;58:539-547. doi: 10.1007/s00592-020-01623-4 [DOI] [PubMed] [Google Scholar]
- 7. Provenzano V, Guastamacchia E, Brancato D, et al. Closing the loop with OpenAPS in people with type 1 diabetes—experience from Italy. Diabetes. 2018;67(suppl 1). https://diabetes.diabetesjournals.org/content/67/Supplement_1/993-P [Google Scholar]
- 8. Choi SB, Hong ES, Noh YH. Open artificial pancreas system reduced hypoglycemia and improved glycemic control in patients with type 1 diabetes. Diabetes. 2018;67(suppl 1). https://diabetes.diabetesjournals.org/content/67/Supplement_1/964-P [Google Scholar]
- 9. Zabinsky J, Howell H, Ghezavati A, Lewis DM, Nguyen A, Wong J. Do-it-yourself artificial pancreas systems for type 1 diabetes reduce hyperglycemia without increasing hypoglycemia. Diabetes. 2020;69(suppl 1). https://diabetes.diabetesjournals.org/content/69/Supplement_1/988-P [Google Scholar]
- 10. Wu Z, Luo S, Zheng X, et al. Use of a do-it-yourself artificial pancreas system is associated with better glucose management and higher quality of life among adults with type 1 diabetes. Ther Adv Endocrinol Metab. 2020;11:2042018820950146. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11. Koutsovasilis A, Sotiropoulos A, Pappa M, et al. Clinical Evaluation of a Closed-Loop Insulin Delivery System on Glycaemic Control in Adults with Type 1 Diabetes. EASD; 2019. Accessed May 14, 2021. https://www.easd.org/virtualmeeting/home.html#!resources/clinical-evaluation-of-a-closed-loop-insulin-delivery-system-on-glycaemic-control-in-adults-with-type-1-diabetes-134c4ad2-53f1-4698-927f-5bd287a76a5d. [Google Scholar]
- 12. Petruzelkova L, Soupal J, Plasova V, et al. Excellent glycemic control maintained by open-source hybrid closed-loop AndroidAPS during and after sustained physical activity. Diabetes Technol Ther. 2018;20:744-750. doi: 10.1089/dia.2018.0214 [DOI] [PubMed] [Google Scholar]
- 13. Toffanin C, Kozak M, Sumnik Z, et al. In silico trials of an open-source android-based artificial pancreas: a new paradigm to test safety and efficacy of do-it-yourself systems. Diabetes Technol Ther. 2020;22:112-120. doi: 10.1089/dia.2019.0375 [DOI] [PubMed] [Google Scholar]
- 14. Burnside M, Lewis D, Crocket H, et al. CREATE (Community deRivEd AutomaTEd insulin delivery) trial. Randomised parallel arm open label clinical trial comparing automated insulin delivery using a mobile controller (AnyDANA-loop) with an open-source algorithm with sensor augmented pump therapy in type 1 diabetes. J Diabetes Metab Disord. 2020;19:1-15. doi: 10.1007/s40200-020-00547-8 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15. Leelarathna L, Choudhary P, Wilmot EG, et al. Hybrid closed-loop therapy: where are we in 2021? Diabetes Obes Metab. 2021;23:655-660. [DOI] [PubMed] [Google Scholar]
- 16. Brogi E, Cyr S, Kazan R, Giunta F, Hemmerling TM. Clinical performance and safety of closed-loop systems: a systematic review and meta-analysis of randomized controlled trials. Anesth Analg. 2017;124(2):446-455. doi: 10.1213/ANE.0000000000001372 [DOI] [PubMed] [Google Scholar]
- 17. Bekiari E, Kitsios K, Thabit H, et al. Artificial pancreas treatment for outpatients with type 1 diabetes: systematic review and meta-analysis. BMJ. 2018;361:k1310. doi: 10.1136/bmj.k1310 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18. Grant AD, Lewis DM, Kriegsfeld LJ. Multi-timescale rhythmicity of blood glucose and insulin delivery reveals key advantages of hybrid closed loop therapy. J Diabetes Sci Technol. Published online March 10, 2021. doi: 10.1177/1932296821994825 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19. Lewis D. Setting expectations for successful artificial pancreas/hybrid closed loop/automated insulin delivery adoption. J Diabetes Sci Technol. 2018;12(2):533-534. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Iturralde E, Tanenbaum ML, Hanes SJ, et al. Expectations and attitudes of individuals with type 1 diabetes after using a hybrid closed loop system. Diabetes Educ. 2017;43(2):223-232. doi: 10.1177/0145721717697244 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Boughton CK, Hartnell S, Allen JM, Fuchs J, Hovorka R. Training and support for hybrid closed-loop therapy. J Diabetes Sci Technol. Published online September 11, 2020. doi: 10.1177/1932296820955168 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22. OpenAPS Reference Design. Principles of an open artificial pancreas system (OpenAPS). Accessed March 30, 2021. https://openaps.org/reference-design/
- 23. oref1. Accessed March 30, 2021. https://openaps.readthedocs.io/en/latest/docs/Customize-Iterate/oref1.html
- 24. Lal RA, Ekhlaspour L, Hood K, Buckingham B. Realizing a closed-loop (artificial pancreas) system for the treatment of type 1 diabetes. Endocr Rev. 2019;40(6):1521-1546. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25. Lewis D. Automated Insulin Delivery: how artificial pancreas “closed loop” systems can aid you in living with diabetes. 2019. https://www.amazon.com/Automated-Insulin-Delivery-artificial-pancreas/dp/1797763695/
- 26. Lewis DM, Leibrand S. Automatic estimation of basals, ISF, and carb ratio for sensor-augmented pump and hybrid closed-loop therapy. Diabetes. 2016;65(suppl 1). https://www.abstractsonline.com/pp8/#!/4297/presentation/45926 [Google Scholar]
- 27. Lewis D. Automated insulin delivery in real life (AID-IRL): real-world user perspectives on commercial AID. J Diabetes Sci Technol. Published online September 16, 2020. doi:10.1177/1932296820957360. https://www.abstractsonline.com/pp8/#!/4297/presentation/45926 [DOI] [PMC free article] [PubMed]