Skip to main content
. 2021 Nov 24;6(48):32637–32646. doi: 10.1021/acsomega.1c04266

Table 1. MS/MS Conditions for Quantification of Rottlerin in Plasma.

parameter value
scan type SRM
source H-ESI
ion polarity negative
vaporizer temperature (°C) 300
ion transfer tube temperature (°C) 250
sheath gas (arbitrary scale) 30
auxiliary gas (arbitrary scale) 10
CID gas (mTorr) 1.5
dwell time per transition (ms) 148
RF lens for rottlerin (V) 159
RF lens for IS (V) 125
Q1 resolution (FWHM) 0.7
Q3 resolution (FWHM) 0.7
collision energy for rottlerin (V) 24
collision energy for IS (V) 20
ion transition for rottlerin (m/z) 515.2 → 321.2
ion transition for IS (m/z) 168.0→ 132.1