. 2021 Nov 26;10(23):5546. doi: 10.3390/jcm10235546

Table 2.

Geometric deviations between the two imaging modalities (CBCT and MRI) were compared for all possible inter-rater combinations separately, for the implant tip, entry-level, and axis. Significant differences are marked in bold. * To avoid artificially increasing the study power, no statistical analysis of mean values was performed.

Parameter Rater Combination Mean Difference between Modalities 95% Confidence Interval p -Value
(Bonferroni-Corrected)
Tip [mm] Mean 0.5 (0.3, 0.7) *
1 vs. 2 0.4 (−0.1, 1) 0.033
1 vs. 3 0.2 (−0.3, 0.7) 0.084
1 vs. 4 0.6 (0.1, 1.1) <0.001
2 vs. 3 0.2 (−0.2, 0.6) 0.399
2 vs. 4 0.6 (0.1, 1) 0.669
3 vs. 4 1.0 (0.5, 1.5) 1
Entry-level [mm] Mean 0.4 (0.3, 0.6) *
1 vs. 2 0.5 (0, 0.9) 0.030
1 vs. 3 0.3 (−0.1, 0.7) 0.156
1 vs. 4 0.4 (0, 0.8) <0.001
2 vs. 3 0.4 (0.1, 0.6) 0.147
2 vs. 4 0.5 (0.1, 0.8) 0.060
3 vs. 4 0.7 (0.3, 1) 0.399
Axis [°] Mean 1.4 (0.7, 2.2) *
1 vs. 2 1.2 (−0.8, 3.2) 0.102
1 vs. 3 1.0 (−0.6, 2.6) 0.012
1 vs. 4 2.8 (1, 4.6) 0.039
2 vs. 3 −0.1 (−2.1, 1.8) 0.720
2 vs. 4 1.7 (0.1, 3.3) 1
3 vs. 4 2.2 (0.5, 3.9) 0.675