. 2021 Dec 3;26(23):7358. doi: 10.3390/molecules26237358

Table 2.

Stability data of lyso-Gb3 assay.

Lyso-Gb3 Amount Spiked in Blank Plasma (ng/mL) Stability Test Mean ± SD
(ng/mL)
n = 3
Precision
(%)
Accuracy
(%)
0.50 3 h (short stability) 0.52 ± 0.02 3.0 4.6
24 h (autosampler stability) 0.52 ± 0.03 4.7 5.6
30 days −20 °C (long stability) 0.53 ± 0.02 3.2 7.1
30 days −80 °C (long stability) 0.52 ± 0.04 7.9 3.3
100.00 2 h (short stability) 93.68 ± 2.02 2.2 −6.3
24 h (autosampler stability) 95.93 ± 2.58 2.7 −4.1
30 days −20 °C (long stability) 106.83 ± 4.05 3.8 6.8
30 days −80 °C (long stability) 103.29 ± 8.21 7.9 3.3

Precision (RSD%): 100 × SD/mean; accuracy (%): 100 × (mean concentration found—known concentration)/known concentration); LLOQ: lower limit of quantification.