Table 1.

Method validation data for vanadium (V) in plasma

Parameter Result
Concentration range (ng V/mL plasma) 5.00 – 1,000
Correlation coefficient (r) > 0.99
Matrix standard accuracy (%RE) −1.4 – 5.1%
LODa (ng V/mL plasma) 0.268
LOQb (ng V/mL plasma) 5.00
Selectivity (mean matrix blank response relative to LOQ, %) 17.3%
Mean Instrument drift (%) 16.4%
Absolute recovery (%) 110 – 122%
Precision and accuracyc
 Intra-day accuracy (mean %RE) 8.5 – 15.6%
 Intra-day precision (%RSD)d ≤ 1.8%
 Inter-day accuracy (mean %RE) 7.3 – 11.7%
 Inter-day precision (%RSD) ≤ 3.1%
Dilution verification (up to 7,500 ng V/mL plasma)
 Accuracy (mean %RE) 5.6%
 Precision (%RSD) 0.9%
a

LOD = Limit of detection.

b

LOQ = Lower limit of quantitation, or the lowest calibration standard that exhibited −20% to 20% relative error (RE).

c

Precision and accuracy were determined for triplicate QC samples fortified at three levels, 7.50, 250, and 800 ng V/mL plasma.

d

%RSD = percent relative standard deviation.