Skip to main content
. 2021 Dec 10;525:46–53. doi: 10.1016/j.cca.2021.12.004

Table 3.

Coincidence rate of the five commercial RT-qPCR diagnostic test kits for SARS-CoV-2 evaluated in this study.

ID Da An
Liferiver
Kinghawk
eDiagnosis
Maccura
ORF1ab N ORF1ab N E ORF1ab N ORF1ab N ORF1ab N E
P1 35.46 34.41 36.17 34.22 32.45 36.28 37.28 36.13 33.71 33.5 33.49 31.8
P2 36.49 35.91 37.37 36.26 37.08 37.57 37.07 38.16 35.9 35.61 36.79 33.21
P3 29.44 27.87 30.72 28.5 26.8 31.63 30.6 30.69 28.43 27.7 25.76 28.05
P4 33.2 31.56 34.73 32.3 30.82 34.91 34.36 34.14 32 31.75 29.28 31.8
P5 31.37 29.89 32.92 30.87 29.2 33.88 32.76 32.88 30.71 30.22 30.36 27.99
P6 36.81 34.34 36.42 35.14 32.95 37.19 No ct 37.58 34.72 34.03 34.19 31.78
P7 31.71 30.03 32.54 30.7 28.79 33.81 32.34 32.06 30.09 29.04 30.63 27.34
P8 26.52 24.93 27.71 25.87 23.73 29.67 27.3 27.21 25.22 23.90 22.45 25.51
P9 28.95 26.84 29.99 27.99 25.62 30.97 29.58 29.61 27.33 26.61 26.85 24.36
P10 34.26 32.05 35.47 33.44 30.98 35.45 34.72 34.41 32.30 31.04 32.16 29.26
P11 No ct No ct No ct No ct No ct No ct No ct No ct No ct No ct No ct No ct
P12 No ct No ct No ct No ct No ct No ct No ct No ct No ct No ct No ct No ct
P13 No ct No ct No ct No ct No ct No ct No ct No ct No ct No ct No ct No ct
P14 No ct No ct No ct No ct No ct No ct No ct No ct No ct No ct No ct No ct
P15 No ct No ct No ct No ct No ct No ct No ct No ct No ct No ct No ct No ct
P16 No ct No ct No ct No ct No ct No ct No ct No ct No ct No ct No ct No ct
P17 No ct No ct No ct No ct No ct No ct No ct No ct No ct No ct No ct No ct
P18 No ct No ct No ct No ct No ct No ct No ct No ct No ct No ct No ct No ct
P19 No ct No ct No ct No ct No ct No ct No ct No ct No ct No ct No ct No ct
P20 No ct No ct No ct No ct No ct No ct No ct No ct No ct No ct No ct No ct
Positive mean 32.42 30.78 33.40 31.53 29.84 34.14 32.89 33.29 31.04 30.34 30.20 29.11
Positive SD 3.44 3.56 3.18 3.36 3.90 2.68 3.35 3.51 3.37 3.62 4.30 2.97
Coincidence rate (%) 100 100 100 100 100 100 95 100 100 100 100 100