Table 1.

Comparison of features of the currently available CGM systems.

Dexcom G6 FreeStyle Libre 2 Eversense Medtronic Guardian 3
Non-adjunctive Dosing Approved Approved Approved Not approved
Calibration Not Required, Can perform Not required, Cannot Perform Every 12 hours Every 12 hours
Sensor Warm-Up 2 hours 1 hour ∼ 26 hours Up to 2 hours
Sensor Wear Time 10 days 14 days (sensor/transmitter in one device) 90 days 7 days
Transmitter Wear Time 90 days 1 year 1 year
MARD 9.0% a 9.2% b 8.5% c 8.7%/ 9.1% d
Customizable Alarms for Hypo- / Hyperglycemia Yes Yes Yes Yes
Data Display Dexcom Receiver or Smart Devices FreeStyle Libre or Smart Devices Dexcom Receiver or Smart Devices 770G or Smart Device (Guardian Connect or MiniMed Mobile)
Follow App Dexcom Share LibreLinkUp Eversense NOW Carelink Connect
Interfering Substances- False Highs Acetaminophen (>1 gram every 6 hrs) Hydroxyurea Vitamin C Mannitol (IV, local irrigation, peritoneal dialysis) Acetaminophen (any dose)
Interfering Substances- False Lows Aspirin Tetracycline
AID Integration Tandem Control IQ No No Medtronic 770G

*MARD, Mean Absolute Relative Difference; AID, Automated insulin delivery.

aShah VN, et al. Diabetes Technol Ther. 2018;20(6):428–433.

bFreeStyle Libre 2 User Manual.

cChristiansen MP, et al. Diabetes Technol Ther. 2019;21(5):231–237.

dChristiansen MP, et al. Diabetes Technol Ther. 2017;19(8):446–456.