Table 4. Positive and negative agreement for the detection of SARS-CoV-2 and influenza A/B between the three commercial kits.
Viruses | PowerChek | STANDARD M | Allplex | |
---|---|---|---|---|
SARS-CoV-2 | ||||
PPA | 90/97 | 93/97 | 93/97 | |
92.8 (85.9–96.5) | 95.9 (89.9–98.4) | 95.9 (89.9–98.4) | ||
NPA | 201/201 | 201/201 | 201/201 | |
100.0 (98.1–100.0) | 100.0 (98.1–100.0) | 100.0 (98.1–100.0) | ||
Kappa | 0.95 (0.91–0.99) | 0.97 (0.94–1.00) | 0.97 (0.94–1.00) | |
Influenza A | ||||
PPA | 71/71 | 71/71 | 71/71 | |
100.0 (94.9–100.0) | 100.0 (94.9–100.0) | 100.0 (94.9–100.0) | ||
NPA | 332/332 | 331/332 | 331/332 | |
100.0 (98.9–100.0) | 99.7 (98.3–100.0) | 99.7 (98.3–100.0) | ||
Kappa | 0.97 (0.93–1.00) | 0.99 (0.98–1.00) | 0.99 (0.98–1.00) | |
Influenza B | ||||
PPA | 50/50 | 50/50 | 50/50 | |
100.0 (92.9–100.0) | 100.0 (92.9–100.0) | 100.0 (92.9–100.0) | ||
NPA | 353/353 | 353/353 | 353/353 | |
100.0 (98.9–100.0) | 100.0 (98.9–100.0) | 100.0 (98.9–100.0) | ||
Kappa | 1.00 (1.00–1.00) | 1.00 (1.00–1.00) | 1.00 (1.00–1.00) |
PPA and NPA were presented as % with 95% confidence intervals and Kappa with 95% confidence intervals.
SARS-CoV-2 = severe acute respiratory syndrome-coronavirus-2, PPA = positive percent agreement, NPA = negative percent agreement.