Skip to main content
. 2021 Nov 30;8:758548. doi: 10.3389/fcvm.2021.758548

Table 4.

Selected cut-off values of the different PGs for predicting clinical relevant OSA based on home measurements.

Sensitivity (%) Specificity (%) PPV (%) NPV (%) Youden's J Index Accuracy
ApneaLink Air (n = 72)
AHI = 7.0 events/h 89.6 58.3 81.1 73.7 0.479 0.79
AHI = 19.9 events/h 52.1 95.8 96.2 50.0 0.479 0.67
Somnotouch RESP (n = 73)
AHI = 10.3 events/h 90.2 50.0 80.7 68.8 0.402 0.78
AHI = 19.9 events/h 56.9 86.4 90.6 46.3 0.433 0.66
SpiderSAS (n = 79)
AHI = 19.5 events/h 87.3 58.3 82.8 66.7 0.460 0.78
AHI = 29.5 events/h 56.4 91.7 93.9 47.8 0.480 0.67

AHI, apnea-hypopnea index; PPV, positive predictive value; NPV, negative predictive value.