Figure 1.

Figure 1.

Schematic diagrams of SRM/MRM, PRM, DDA, and DIA proteomics. The SRM/MRM method isolates and fragments pre-defined precursor ions and monitors pre-defined product ions derived from the precursor. PRM isolates and fragments pre-defined precursors in the same way as SRM/MRM but records all product ions of a pre-defined precursor. DDA automatically selects the most abundant precursor ions for isolation and fragmentation and monitors all product ions. DIA isolates precursors into pre-defined isolation mass windows and fragments all precursors in each window to generate product ions being monitored by an HRAM mass analyzer.