Skip to main content
. 2021 Jan 15;23(4):793–801. doi: 10.1111/jch.14179

TABLE 1.

Comparison of blood pressure parameters measured by the wrist‐type and upper arm‐type devices in the office setting (100 pairs of measurements from 50 patients)

Wrist device Upper arm device Difference (Wrist—upper arm) p for difference
Sitting position
SBP, mmHg 129.7 ± 16.1 130.2 ± 14.1 −0.5 ± 9.7 .607
DBP, mmHg 76.5 ± 12.1 78.9 ± 10.4 −2.4 ± 5.9 <.001
Heart rate, bpm 72.0 ± 12.1 72.1 ± 11.8 −0.1 ± 4.1 .753
Supine position
SBP, mmHg 129.6 ± 14.9 127.8 ± 12.4 1.8 ± 8.5 .042
DBP, mmHg 74.9 ± 10.0 77.9 ± 8.9 −3.0 ± 5.4 <.001
Heart rate, bpm 67.4 ± 10.6 67.5 ± 10.5 −0.1 ± 2.2 .655

Values are mean ± SD.

Abbreviations: DBP, diastolic blood pressure; SBP, systolic blood pressure.