Skip to main content
. 2000 Jun;38(6):2416–2418. doi: 10.1128/jcm.38.6.2416-2418.2000

TABLE 2.

Summary of improved reproducibility following procedural changes for the Abbott LCx C. trachomatis and N. gonorrhoeae tests

Parameter Results:
Before changes After changes
No. of samples 14,420 7,728
No. of repeats 181 514a
No. of discrepant results/no. of repeats (%) 30/181 (17) 13/514 (2.5)
No. of discrepant results detected by the Abbott equivocal range/total no. of discrepant results (%) 5/30 (17) 6/13 (46)
Mean difference (S/CO ratio [n]) between pairs of results
 Urine 2.0 (14) 1.9 (2)
 Swabs 0.88 (16) 0.49 (11)b
a

Includes 475 urine specimens that were tested in duplicate. 

b

Differs significantly (P < 0.05) from values before changes.