Table 1. The limit of detection for the UC Santa Cruz Multiplex SARS-CoV-2 assay.
Dilution (vg/uL) | N1 positive | N2 positive | Cq Mean N1 | Cq Mean N2 | Cq Mean RNase P |
---|---|---|---|---|---|
10 | 3/3 | 3/3 | 31.66 | 30.89 | 25.02 |
5 | 3/3 | 3/3 | 32.69 | 31.84 | 25.10 |
2.5 | 3/3 | 3/3 | 33.98 | 33.33 | 25.11 |
1.25 | 3/3 | 3/3 | 34.25 | 33.50 | 24.91 |
0.625 | 3/3 | 3/3 | 35.18 | 34.63 | 24.99 |
0.3125 | 2/3 | 2/3 | 36.46 | 35.41 | 24.99 |
0.156 | 1/3 | 1/3 | 37.33 | 36.89 | 24.65 |
0.078 | 0/3 | 0/3 | ND | ND | 24.39 |
0.039 | 1/3 | 1/3 | 37.18 | 35.90 | 24.93 |
0 | 0/3 | 0/3 | ND | ND | 24.65 |