Table 3. Validation of UC Santa Cruz Multiplex SARS-CoV-2 assay by a commercial lab using true positive and true negative asymptomatic surveillance samples.
Samples Tested Individually | Comparator Method Result | |
---|---|---|
Candidate Test Result | Positive | Negative |
Positive | 29 | 0 |
Negative | 1 | 30 |
(samples collected on asymptomatic patients).
Performance of the UCSC Multiplex SARS-CoV-2 Assay against the Pangea Comparator: Positive Percent Agreement: 29/30 = 97% (95%CI: 84%-100%), Negative Percent Agreement: 30/30 = 100% (95%CI: 90%-100%).