Dose rates recorded in the central region of the UVC cabinet measured in different positions at different heights from the bottom of the cabinet

Position inside the cabinet Without aluminum foil With aluminum foil
Dose rate (mW/cm2) Dose rate (mW/cm2)
0.0 cm from the base 0.57 ± 0.19 8.5 ± 0.5
6.5 cm from the base 3.6 ± 0.7 8.6 ± 0.5
12.0 cm from the base 2.3 ± 0.5 8.4 ± 0.5
18.5 cm from the base 3.0 ± 0.6 8.2 ± 1.0
27.0 cm from the base 4.0 ± 0.6 10.2 ± 0.5
36.0 cm from the base 3.0 ± 0.5 7.5 ± 0.2
Cabinet average 2.7 ± 0.5 8.6 ± 0.3