Table 2.
Sensitivity and specificity of SARS-CoV-2-N-IgG ELISA with reference to the Colloidal gold antibody detection kit
Colloidal gold antibody detection kit | r-SARS-CoV-2-N based IgG ELISA | Total | |
---|---|---|---|
Positive Negative | |||
Positive | 174 | 0 | 174 |
Negative | 6 | 182 | 188 |
Total | 180 | 182 | 362 |
Concordance a: 98.3% | Sensitivity b: 100% | Specificity c: 96.8% |
χ2 = 3.702, p > 0.05
The sensitivity and specificity of the SARS-CoV-2-N IgG ELISA compared to the Colloidal gold antibody detection kit
a) (The number of positive samples detected by both methods + the number of samples determined to be negative by both methods)/the total number × 100
b) True positive (true positive + false negative) ×100
c) True negative (true negative – false positive) × 100