Table 1.
Statistical evaluation of calculated concentrations of FGAR obtained from the calibration curves (n=3) prepared in mouse plasma and mouse tumor homogenates.
| Mouse Plasma | Mouse Tumor Homogenate | ||||
|---|---|---|---|---|---|
| Nominal Conc. (nmol/mL) | Accuracy (%) | Precision (% RSD) | Nominal Conc. (nmol/g) | Accuracy (%) | Precision (% RSD) |
| 100 | 94.4 | 5.18 | 10000 | 98.9 | 0.46 |
| 32.0 | 103 | 8.64 | 3000 | 102 | 1.21 |
| 10.0 | 98.9 | 6.76 | 1000 | 102 | 1.51 |
| 3.20 | 105 | 5.31 | 300 | 101 | 1.74 |
| 1.00 | 98.6 | 4.63 | 100 | 100 | 3.27 |
| 0.32 | 97.1 | 10.3 | 30 | 104 | 1.32 |
| 0.10 | 100 | 9.05 | 10 | 95.9 | 1.31 |
| 0.03 | 101 | 7.10 | 3 | 93.0 | 4.83 |
| - | - | - | 1 | 103 | 1.52 |