Skip to main content
. 2021 Jun 10;107(1):20–25. doi: 10.1136/archdischild-2020-321387

Table 2.

Proportions of time within SpO2 ranges

Oxygenie CLiO2 P value*
Time SpO2 within target range† 80.2 (72.6–82.4)% 68.5 (56.7–79.3)% 0.005
Time SpO2 below target range 14.7 (11.8–17.2)% 9.3 (8.2–12.6)% 0.020
Time SpO2 above target range‡ 6.3 (5.1–9.9)% 15.9 (11.5–30.7)% 0.001
SpO2 85%–90% 12.6 (10.9–13.1)% 8.5 (7.6–11.0)% 0.020
SpO2 80%–84% 1.2 (0.7–3.0)% 0.8 (0.5–0.9)% 0.003
SpO2 <80% 0.5 (0.1–1.0)% 0.2 (0.1–0.4)% 0.061
SpO2 96%–98% while FiO2≥0.22 6.1 (5.0–9.5)% 15.5 (10.9–27.4)% 0.001
SpO2 >98% while FiO2≥0.22 0.2 (0.1–0.4)% 1.4 (0.4–3.7)% 0.001

Data in median (IQR)

*Wilcoxon matched pairs test.

†91% ≤SpO2 ≤95% or SpO2 ≥96% while FiO2=0.21.

‡SpO2 ≥96% while FiO2 ≥0.22.