TABLE 2.
Previous IPG (cDBS mode) |
AlphaDBSipg (cDBS mode) |
|||||||||
Patient | Stim Config | Imp (kΩ) | Amp (mA) | Freq (Hz) | PW (μs) | TEED** (μW) | Amp (mA) | Freq (Hz) | PW (μs) | TEED** (μW) |
01 | Left: C + 1− | 0.84 | 4* | 130 | 90 | 187 | 4 | 130 | 60 | 124 |
Right: C + 9− | 0.94 | 3.4* | 130 | 90 | 143 | 3.5 | 130 | 60 | 90 | |
02 | Left: C + 1− | 0.81 | 3.2* | 130 | 60 | 80 | 2.4 | 130 | 60 | 45 |
Right: C + 9− | 0.86 | 2.4* | 130 | 60 | 45 | 1.8 | 130 | 60 | 25 | |
03 | Left: C + 3− | 1.47 | 2.5 | 130 | 60 | 49 | 3.0 | 130 | 60 | 70 |
Right: C + 11− | 0.96 | 2.3 | 130 | 60 | 41 | 2.5 | 130 | 60 | 49 |
*Previous implant voltage controlled. Current values were calculated using a simple translation to current on the basis of measured impedances parameter following Ohm’s law.
**TEED calculated using nominal 1 kΩ impedance.