Skip to main content
. 2021 Dec 8;9:765516. doi: 10.3389/fbioe.2021.765516

TABLE 1.

Assumptions for the uncertainty quantification calculations. U stands for uniform distribution. We distinguish between geometrical and handling uncertainties. The geometrical uncertainties are due to manufacturing inaccuracies or the limited knowledge of the exact geometry. We estimated the geometrical uncertainties of the electrodes based on a measurement of the chamber used in this work. In contrast, the handling uncertainties are introduced by the experimenter.

Parameter Distribution Reasoning
Geometrical uncertainties
 Height of electrode h 1 or h 2/mm U(0.01,1.5) Misalignment of electrodes
 Length of bottom part l b/mm U(21,22.3) Misshaping of electrodes
 Spacing of electrodes d/mm U(23,25) Misalignment of electrodes
 Decay of meniscus profile c/mm U(1.95,2.05) Estimate based on Schuderer and Kuster (2003)
 Height of meniscus profile h 0/mm U(1.8,2.5) Estimate based on Schuderer and Kuster (2003)
Handling uncertainties
 Cell culture medium V/ml U(3.4,3.6) Pipetting inaccuracies