Skip to main content
. 2021 Feb 3;11(11):5896–5902. doi: 10.1039/d0ra09751a

Results from the analytical methods validation.

Parameter Result
Linearity 0.1–3 μg mL−1
Correlation coefficient r = 0.9982
Limit of quantification 0.1 μg mL−1
Limit of detection 0.03 μg mL−1
Stability
Short term 87.56–100.2%
Freeze and thaw cycles 93.45–103.7%
Autosampler 95.92–100.7%
Precision (CV%)
Intraday 7.1
Interday 9.0
Accuracy (RE%)
Intraday 4.4
Interday 6.5
Recovery (%) 87