Skip to main content
. 2000 Jul;38(7):2480–2483. doi: 10.1128/jcm.38.7.2480-2483.2000

TABLE 2.

Discrepant results obtained for cervical swabs tested individually and in pools of four and eight by LCx

Specimen no. Test result (sample rate/cutoff rate)a for:
Individual test Pool of 4 Pool of 8
BP-8765 POS (1,011.1/426.1) POS (583.5/426.1) NEG (252.7/426.1)
46-384519 NEG (184.4/577.1) POS (539.6/462.2)b POS (895.1/577.1) POS (893.1/577.1)
BP-8904 NEG (463.9/577.1) POS (672.7/462.2)b POS (895.1/577.1) POS (893.7/577.1)
BP-9082 POS (660.9/448.3) NEG (11.1/448.3) NEG (160.7/448.3)
BP-16300 POS (2,043.0/654.6) NEG (13.7/654.6) NEG (14.0/654.6)
a

The rate was calculated as described in the manufacturer's package insert (as counts per second per second) and printed from the LCx analyzer. POS, positive; NEG, negative. 

b

The sample was retested after 24 h at 4°C.