Skip to main content
. Author manuscript; available in PMC: 2021 Dec 23.
Published in final edited form as: J Neural Eng. 2020 Oct 8;17(5):056009. doi: 10.1088/1741-2552/abb7a6

Figure 12.

Figure 12.

Mean extent of particulate Pt occupying the UB turn scala tympani as a function of charge density. There was a statistically significant increase in Pt corrosion products associated with charge density (p < 0.01). The charge recovery technique also appeared to effect corrosion product (p < 0.01), with reduced levels associated with AP and CC. Error bars = sem.