The characteristics of the response of the two sensors towards pioglitazone and the validation parametersa.
| Parameters | SC-SP-ISE | LC-ISE | HPLC | ||
|---|---|---|---|---|---|
| Analytes | PIO | PIO | PIO | ALG | |
| Slope | 59.40 | 57.20 | 864.64 | 1464.70 | |
| Intercept | 663.80 | 604.80 | −33.57 | 188.20 | |
| The correlation coefficient (R2) | 0.9990 | 0.9990 | 0.9999 | 0.9998 | |
| The linear range (molar) | 1 × 10−6 to 1 × 10−2 | 1 × 10−5 to 1 × 10−2 | 2.55 × 10−5 to 20.36 × 10−5 | 1.27 × 10−5 to 10.18 × 10−5 | |
| Working pH range | 1–5.5 | 1–5.5 | — | — | |
| Response time (s) | 10 | 15 | — | — | |
| Retention time (min) | — | — | 2.824 | 1.179 | |
| Validation parameters | |||||
| Accuracy (mean recovery ± SD) | 99.27 ± 0.38 | 99.67 ± 0.78 | 100.07 ± 0.48 | 100.13 ± 0.95 | |
| Repeatability (RSD%) | 0.66 | 0.73 | 0.75 | 0.55 | |
| Intermediate precision (RSD%) | 0.61 | 0.52 | 0.98 | 0.47 | |
| LOD | 2.4 × 10−7 M | 1.7 × 10−6 M | |||
| Robustness* | a | — | — | 100.05 ± 0.80 | 99.94 ± 0.32 |
| b | 99.15 ± 0.58 | 100.24 ± 0.91 | |||
| c | 100.87 ± 0.63 | 99.84 ± 0.72 | |||
| d | 100.61 ± 0.78 | 99.08 ± 0.23 | |||
* is the average recoveries of three determinations of both PIO and ALG using: (a) mobile phase at pH = 2.8, instead of 3; (b) mobile phase contains 67% methanol, instead of 70%; (c) flow rate at 1.15, instead of 1.35; and (d) column temperature at 28 instead of 30 °C.