Quality assurance of the proposed method, comparing with the method reported by Amer et al. (2013).
| Compound | Validation according to the Amer et al. (2013) method49 | This work | ||||
|---|---|---|---|---|---|---|
| Linear range (μg L−1) | R 2 | LOD (μg L−1) | Linear range (μg L−1) | R 2 | LOD (μg L−1) | |
| PAME | 500.0–50 000 | 0.992 | 166.0 | 0.03–500 | 0.997 | 0.06 |
| SAME | 500.0–50 000 | 0.995 | 200.0 | 0.03–500 | 0.996 | 0.06 |
| OAME | 500.0–50 000 | 0.993 | 250.0 | 0.03–500 | 0.996 | 0.03 |
| LAME | 500.0–50 000 | 0.995 | 250.0 | 0.03–500 | 0.997 | 0.05 |