Skip to main content
. 2021 Dec 20;10(12):3160. doi: 10.3390/foods10123160

Table 1.

Validation data for the optimized analytical method using HILIC-CAD, including concentration range, linear equation, correlation coefficients, LOD, LOQ, intra- and inter-day precision, and accuracy.

Saccharides Fructose Glucose Sucrose 1-Kestose Nystose Fructofuranosyl
Nystose
Concentration range
(µg/mL)
0.5–100 0.5−100 0.5−100 0.5−100 0.5−100 0.5−100
Linear equation y = −0.0003x2 + 0.1193x + 0.0279 y = −0.0003x2 + 0.1268x + 0.0577 y = −0.0003x2 + 0.1237x + 0.0781 y = −0.0003x2 + 0.1149x + 0.0539 y = −0.0002x2 + 0.0980x + 0.0329 y = −0.0002x2 + 0.0995x + 0.0241
R2 0.9998 0.9998 0.9996 0.9998 0.9997 0.9999
LOD/LOQ
(µg/mL)
0.25/0.5 0.25/0.5 0.25/0.5 0.25/0.5 0.25/0.5 0.25/0.5
Precision (%) Intra-day
(n = 9)
0.733 0.768 0.639 0.668 0.848 0.840
Inter-day
(n = 3)
0.630 1.46 1.02 1.67 1.46 1.71
Accuracy
(%)
96.5 91.7 94.3 92.3 98.2 97.9