Defibrillation efficacy analysis on upper torso concentrated sections, M1 = Inline graphicmyo > 30V/cm, M2 = Inline graphicmyo > 45V/cm, M3 = Inline graphicmyo > 60V/cm, (loc-location, O-original, E-Energy, J-Joule, the units of D1, D2, andD3 are meter (m) and A1 is m2).

loc D1 D2 D3 A1 DFT(V) E(J) WKLD M1 M2 M3
p1 0.066 0.052 0.056 0.009 268.5 4.687 8.099 6.092 0.868 0.129
p2 0.042 0.030 0.067 0.008 269.9 4.734 10.535 6.568 1.007 0.138
p3 0.066 0.052 0.056 0.009 272.0 4.809 6.204 4.756 0.564 0.102
p4 0.083 0.075 0.053 0.009 326.9 6.944 5.340 4.584 0.576 0.111
p5 0.098 0.102 0.053 0.009 387.3 9.750 5.721 5.109 0.612 0.102
a1 0.040 0.052 0.022 0.008 159.0 1.642 3.260 1.556 0.252 0.0667
a2 0.046 0.075 0.029 0.008 205.3 2.740 4.792 3.177 0.646 0.182
a3 0.051 0.102 0.032 0.009 255.2 4.234 3.745 2.232 0.611 0.249
a4 0.046 0.075 0.029 0.008 184.4 2.209 4.095 2.643 0.405 0.118
a5 0.040 0.052 0.022 0.008 173.1 1.946 3.798 2.174 0.342 0.104