Table 3.

Comparison between the second (Excalibur) and third (X3)-generation balloon procedures characteristics.

Excalibur
N = 26
X3
N = 24
p
Ablation
Laser duration on left pulmonary veins—min 33 (21.5–37) 11.6 (9–17.5) <0.001
Laser duration on right pulmonary veins—min 35 (27–42) 9 (6.6–13.5) <0.001
Total Laser duration—min 69 (55–76) 22 (19–29) <0.001
Number of left veins manual shots 49 (39–55.8) 8 (0–13) <0.001
Number of right veins manual shots 51 (41–58) 0 (0–6) <0.001
Number of manual shots 101 (89–109.8) 14 (8.3–21.3) <0.001
Number of left veins RAPID shots 0 2.5 (0–6)
Number of right veins RAPID shots 0 3 (0–5.8)
Number of RAPID shots 0 12 (0–19.5)
All PV isolated at first pass 15 (58) 17 (71) 0.388
All PV isolated at procedure end 24 (92) 22 (92) 1
AF onset during procedure 3 (12) 2 (8) 1
Procedure data
Procedure duration—min 151.5 (128.5–167) 111 (100–128) <0.001
Iodinated contrast media—ml 2 (2–3) 2 (2–3.8) 0.501
Fluroscopy duration—min 17.8 (12.9–23.3) 14.9 (11.5–24.4) 0.382
Dose-area product—cGy.cm2 443.5 (268.3–775.5) 439 (220–869) 0.645
Complication—no (%) 1 (4) 1 (4)
Cardiac tamponade 1 (4 0
Transient phrenic nerve palsy 0 1 (4)
AF recurrence—no (%) 4 (15) 3 (13) 1

AF: atrial fibrillation; PV: pulmonary veins.