Skip to main content
. 2021 Dec 7;21(24):8179. doi: 10.3390/s21248179

Table 6.

Accuracy analysis of water temperature, pH, DO, and EC, and salinity levels for in situ measurement.

Items Accuracy Analysis
Mean Difference Range MAE RMSE
Temperature sensor 25.0951 °C −0.31–0.34 °C 0.0961 °C 0.1178 °C
pH sensor 7.1755 −0.20–0.20 0.0640 0.782
DO sensor 9.5228 mg/L −0.35–0.33 mg/L 0.0990 mg/L 0.1217 mg/L
EC sensor 0.7923 mS/cm −0.18–0.18 mS/cm 0.0571 mS/cm 0.0694 mS/cm
Salinity estimation 464 ppm −2–2 ppm 0.5337 ppm 0.7557 ppm