. 2021 Dec 7;21(24):8179. doi: 10.3390/s21248179

Table 7.

Accuracy analysis of water temperature, pH, DO, and EC, and salinity levels for post-calibration.

Items Reference Accuracy Analysis
Difference Range MAE RMSE
Temperature sensor HEL-711 RTD −0.10–0.10 °C 0.0339 °C 0.0417 °C
pH sensor pH 6.86 buffer solution −0.23–0.15 0.0603 0.0738
DO sensor FOPTOD ODO −0.10–0.10 mg/L 0.0329 mg/L 0.401 mg/L
EC sensor 1413 μS/cm
conductivity solution
−0.19–0.17 mS/cm 0.0574 mS/cm 0.0698 mS/cm
Salinity estimation Vernier Salinity Sensor −10–10 ppm 2.9743 ppm 3.6078 ppm