Abstract
Background
Rapid and irregular ventricular rates (RVR) are an important consequence of atrial fibrillation (AF). Raw accelerometry data in combination with electrocardiogram (ECG) data have the potential to distinguish inappropriate from appropriate tachycardia in AF. This can allow for the development of a just-in-time intervention for clinical treatments of AF events. The objective of this study is to develop a machine learning algorithm that can distinguish episodes of AF with RVR that are associated with low levels of activity.
Methods
This study involves 45 patients with persistent or paroxysmal AF. The ECG and accelerometer data were recorded continuously for up to 3 weeks. The prediction of AF episodes with RVR and low activity was achieved using a deterministic probabilistic finite-state automata (DPFA)-based approach. Rapid and irregular ventricular rate (RVR) is defined as having heart rates (HR) greater than 110 beats per minute (BPM) and high activity is defined as greater than 0.75 quantile of the activity level. The AF events were annotated using the FDA-cleared BeatLogic algorithm. Various time intervals prior to the events were used to determine the longest prediction intervals for predicting AF with RVR episodes associated with low levels of activity.
Results
Among the 961 annotated AF events, 292 met the criterion for RVR episode. There were 176 and 116 episodes with low and high activity levels respectively. Out of the 961 AF episodes, 770 (80.1%) were used in the training data set and the remaining 191 intervals were held out for testing. The model was able to predict AF with RVR and low activity up to 4.5 min before the events. The mean prediction performance gradually decreased as the time to events increased. The overall Area under the ROC Curve (AUC) for the model lies within the range of 0.67–0.78.
Conclusion
The DPFA algorithm can predict AF with RVR associated with low levels of activity up to 4.5 min before the onset of the event. This would enable the development of just-in-time interventions that could reduce the morbidity and mortality associated with AF and other similar arrhythmias.
Keywords: Machine learning, Atrial fibrillation, Arrhythmia prediction, Signal processing, Probabilistic finite-state automata
Introduction
are an important consequence of [1]. A fast and irregular heart rate decreases time spent in diastole which impairs myocardial perfusion, ventricular diastolic filling, and cardiac output that in turn leads to greater symptom burden and reduced left ventricular systolic function [2, 3]. However, there are many reasons for RVR during episodes of AF. In addition to inappropriate tachycardia due to the underlying disease, patients’ heart rates may rise in a normal compensatory response to keep up with physiological demands during exercise or activities of daily living.
There is a need to not only detect and predict periods of AF with rapid ventricular rates but also distinguish periods of inappropriate tachycardia with AF from those associated with activity. Accelerometers are widely used in wearable sensors and provide important data that can be used to estimate patients’ activity and movement within their natural setting [4]. They can provide important insights that are not available through investigating alone. Therefore using raw accelerometry data in combination with ECG has the potential to identify and predict inappropriate tachycardia with AF. Accurate prediction of these episodes will allow just-in-time interventions for patients with AF.
Traditional algorithms have focused on the detection of AF episodes and are able to achieve this with high degree of accuracy [5]. However, prediction algorithms have not been able to achieve sufficient lag time to allow for medical or surgical interventions. There are several clinical scenarios where identifying rapid AF at low levels of activity can be of clinical value, including ensuring appropriate rate control for patients with AF, triggers for on-demand rhythm control approaches to improve patient care and facilitate monitoring system for AF that could be treated medically.
The aim of this study is to predict the onset of AF with RVR episodes associated with low activity using pre-event ECG signals. The proposed approach involves a DPFA-based algorithm that can predict episodes of AF with RVR that are associated with low levels of activity.
Methods
A total of 45 patients with history of AF who presented to University of Michigan are recruited in the study. All patients wore an event recorder (Preventice solutions Inc) for up to 3 weeks. IRB approved the protocol and written informed consent was obtained. The ECG and accelerometer data were recorded continuously for up to 3 weeks. Thirteen patients are excluded because the length of the ECG is less than 5 min or there are no annotated AF events, 1 additional patient is excluded due to missing accelerometer data. Thirteen patients with AF episodes of duration less 30 s or less than 5 episodes of AF are further excluded. Finally a total of 18 patients are included in the prediction analysis. A flow chart of the patient inclusion/exclusion criteria is demonstrated in Fig. 1. The characteristics of patients are summarized in Table 1. The demographic data is missing for 1 patient.
Table.1.
Variable | All participants (n=44*) | Participants in prediction analysis (n=18) |
---|---|---|
Female | 14 (31.8%) | 5 (27.8%) |
Age | 66.4 (11.7) | 69.1 (7.3) |
BMI | 31.3 (6.1) | 30.9 (5.7) |
Hypertension | 26 (591%) | 11 (61.1%) |
History of stroke | 0 (0%) | 0 (0%) |
Diabetes | 12 (27.3%) | 5 (27.8%) |
Coronary artery disease | 11 (25%) | 5 (27.8%) |
Peripheral vascular disease | 2 (4.5%) | 2 (11.1%) |
Beta blockers | 31 (70.5%) | 13 (72.2%) |
Calcium channel blockers | 14 (31.8%) | 4 (22.2%) |
Antiarrhythmic drugs | 9 (20.5%) | 2 (11.1%) |
Data sources
Accelerometer data
The raw accelerometer data are collected along three orthogonal axes in the device-specific frame of reference. The continuous accelerometer data are collected using a wireless monitoring device that adheres to patient’s chest (Preventice Solutions, Inc) and sampled at 10 Hz. The accelerometer data were up-sampled to 256 Hz to match the ECG sampling rate. For the analysis, the accelerometer magnitude am(t) consisting of the vector magnitude of the accelerometer data at each time point is used in analysis together with the synchronized ECG signal.
1 |
The data is aggregated using which computes the deviation of am(t) from its mean over the corresponding epoch, averaged over the length of the annotated AF signal [6].
2 |
where:
The quantile relative to the measurements from the entire group is designated as low activity and quantile as high activity in order to account for inter-individual differences in activity levels within the study group. Threshold of 0.75 quantile for distinguishing activity levels is an arbitrary choice since there is no well-established guideline for low versus high activity level.
ECG data
ECG is acquired using a single lead event recorder (Preventice Solutions, Inc) and sampled at 256 Hz. ECG data were collected with the same biopatch as the accelerometer. Continuous single channel ECG signals were collected while wearing the device. Arrhythmia classification was performed using the BeatLogic platform that is cleared to identify AF episodes [5]. Our clinician has also reviewed 150 randomly selected AF events annotated by Beatlogic for further validation, out of which 139 (93%) were confirmed to be AF events. A total of 961 AF events were annotated by the algorithm. Most of the participants had less than 50 episodes of AF events throughout the 3 weeks with several participants with higher number of AF events. Figure 2 shows burden of AF for all participant and those included in the prediction algorithm analysis.
In terms of , RVR episodes are defined as having , and non-RVR episodes with BPM. Pre-processing, noise removal, and prediction interval extraction were performed on the synchronized recordings of ECG and accelerometer magnitude signals.
In the signal pre-processing step, a second-order Butterworth band-pass filter was applied with cutoff frequencies of 0.5 and 40 Hz to the raw ECG signal for noise removal. In the next step, a double median filter with orders equal to 0.2 and 0.6 times the sampling frequency was applied to remove baseline wandering. Peak-detection algorithm was then applied to capture the R-peaks in the ECG signals. Noise section of the signals was detected with annotations from BeatLogic platform.
Events that occurred too close to previous events were excluded to avoid overlap of prediction intervals with arrhythmia events. Events that occurred within 8 min of a noisy signal were also excluded to ensure that the prediction interval is outside the noisy signal range. We used the ECG signal to predict episodes of AF with RVR and low activity level versus all other AF episodes.
The gap interval represents the interval in minutes before the event that is used for prediction. The signal interval represents the length of the prediction signals. For example, prediction intervals with min, min are the signals that span from 3 to 1 mins before the annotated events. Figure 3 shows an example of an annotated AF episode and prediction intervals.
DPFA model
A is a computational model which could be used to generate a string of letters from a fixed alphabet. Given a training sample of strings, a could be effectively trained to summarize and reproduce the predominant rhythm patterns within the training data. In previous work, we have applied a novel -based algorithm to predict , a surrogate for AF, and events by analyzing ECG data several minutes prior to the onset of the cardiac event. In most cases the algorithm outperforms other more traditional and well-established approaches. [7].
In this study, we improved our algorithm and tested its ability to work with different types of synchronized physiological signals (i.e synchronized ECG with accelerometer magnitude signals) for arrhythmia events predictions. We used the DPFA algorithm for predicting AF episodes with RVR and low activity level versus other AF episodes using ECG and accelerometer magnitude signals prior to the event.
Each DPFA is generated in two steps (see [7] for details of the algorithm):
First in the symbolization module, one begins with the training dataset which consists of annotated ECG and accelerometer magnitude signals. From these, the algorithm extracts windows that are indicative of imminent events and others that are not. The positive (i.e., AF with RVR with low activity) and negative (i.e., all the other regions) ECG and accelerometer magnitude signals are respectively combined and transformed into probabilistic strings. Here the probabilistic strings consist of an alphabet of nine symbols where correspond to different activity levels and correspond to different ECG morphology types within each window, and the probabilities
3 |
are computed by assigning probabilities to and independently. The individual probabilities and are obtained from the signal values over discrete time windows t via soft-thresholding
4 |
where the soft-thresholding functions are chosen to be piecewise linear functions of the form
5 |
The parameters were all tuned in the training step and set at
6 |
based on a grid search while running the inner loops of the nested cross validation.
Then the DPFA generation module constructs the positive (AF with RVR and low activity) DPFA and the negative (all other AF episodes) DPFA respectively from the positive and negative probabilistic strings, by first constructing the frequency prefix trees and , followed by the largest suffix merging algorithm and . The frequency prefix trees are tree-like automata with initial state whose state space consists of strings in the alphabet with non-zero frequency, and transition function given by concatenation of strings. The largest suffix merging algorithm selects those states of the frequency prefix tree T with sufficiently high frequency to be in the state space M, and then defines the transition state to be the largest suffix of that is itself contained in the state space of M. See [7] for further details.
The classification scheme contains a training phase and a testing phase. In the training phase, the algorithm learns the DPFA and for the positive and negative classes respectively. Then in the testing phase, we classify a given synchronized ECG-accelerometer magnitude signal from the test dataset by comparing the goodness-of-fit with the DPFA and to predict episodes of AF with RVR with low activity Fig. 4.
Data partition
A total of 961 episodes of AF events were annotated for the analysis. We used 5 fold nested cross validation for parameter tuning. Figure 5 shows how data was partitioned in the analysis. The 961 episodes were split into 5 folds, 4 folds were used for training and the remaining fold was used for testing. Within the inner loop, the training data had a total of 4 folds, 3 folds were used for parameter tuning and the remaining fold for validations. In the outer loop, the DPFA models were generated using all 4 folds of the training data with hyper-parameters tuned during the inner loop. Predictions were performed on the testing data in each split and then averaged.
Results
During the study period, we have recruited 45 patients with AF, of whom 31 had both ECG and accelerometer data. We excluded patients with all AF episodes of duration less than 30 seconds and those with less than 5 episodes of AF. There was a total of 292 episodes of AF with RVR compared to 669 episodes of AF with controlled heart rate. There were 116 episodes in percentile activity level in the RVR group and 124 episodes in the non-RVR group based on MAD relative to total group. The distribution of high activity based on heart rates above 110 BPM are summarized in Table 2. The 45 participants in the study had a median of 22 days of total wear time (range 2–31 days). The median overall AF burden was ( 4.9–86.0%).
Table.2.
Low activity | High activity | Total | |
---|---|---|---|
RVR | 176 | 116 | 292 |
Non-RVR | 545 | 124 | 669 |
Total | 721 | 240 | 961 |
Among the 961 annotated AF events, 292 of them met the criterion for RVR episode, among which 176 episodes had low activity level and the remaining 116 episodes had high activity level based on MAD relative to the entire group activity level. For the rest of the annotated AF events, 669 of them were non-RVR episodes, among which 545 episodes had low activity and the remaining 124 episodes had high activity level. Table 2 summarizes the number of annotated AF episodes by activity and levels.
A total of 18 patients are included in the final DPFA model. They had a median of 22 days of total wear time (range 6–31 days). The median overall AF burden for the 18 patients was ( 10.5–62.3%). Two participants had an AF burden over the duration of the study. The mean heart rate was BPM. The burden of AF for study participants is summarized in Fig. 2.
Different combinations of signal intervals (i.e., min and gap intervals (i.e., min) up to 5 min before the AF events were used for prediction. For a given model based on various gap intervals, the threshold can be adjusted to optimize different parameters, i.e., a more sensitive model versus a more specific model. In Table 3, we report the threshold, sensitivity, specificity, precision, and other relevant data based on each gap interval.
Table.3.
Gap interval (min) | Prediction interval (min) | AUC mean (Std) | Sensitivity mean (Std) | Specificity mean (Std) | Accuracy mean (Std) |
---|---|---|---|---|---|
0.5 | 0.5 | 0.735(0.026) | 0.552(0.049) | 0.808(0.104) | 0.806(0.027) |
1 | 0.5 | 0.700(0.050) | 0.546(0.075) | 0.797(0.085) | 0.811(0.035) |
1.5 | 0.5 | 0.725(0.041) | 0.545(0.098) | 0.804(0.078) | 0.810(0.024) |
2 | 0.5 | 0.706(0.064) | 0.527(0.111) | 0.810(0.084) | 0.799(0.027) |
2.5 | 0.5 | 0.729(0.048) | 0.521(0.103) | 0.845(0.065) | 0.781(0.062) |
3 | 0.5 | 0.703(0.035) | 0.511(0.115) | 0.790(0.091) | 0.758(0.088) |
3.5 | 0.5 | 0.673(0.049) | 0.527(0.121) | 0.742(0.092) | 0.793(0.059) |
4 | 0.5 | 0.701(0.051) | 0.521(0.115) | 0.802(0.110) | 0.718(0.068) |
4.5 | 0.5 | 0.691(0.038) | 0.520(0.145) | 0.790(0.073) | 0.785(0.060) |
0.5 | 1 | 0.768(0.031) | 0.541(0.041) | 0.850(0.081) | 0.779(0.109) |
1 | 1 | 0.756(0.046) | 0.524(0.074) | 0.838(0.097) | 0.841(0.019) |
1.5 | 1 | 0.753(0.019) | 0.545(0.092) | 0.838(0.104) | 0.826(0.021) |
2 | 1 | 0.751(0.054) | 0.488(0.094) | 0.864(0.073) | 0.828(0.043) |
2.5 | 1 | 0.753(0.012) | 0.585(0.102) | 0.805(0.135) | 0.827(0.040) |
3 | 1 | 0.734(0.026) | 0.569(0.116) | 0.767(0.128) | 0.783(0.106) |
3.5 | 1 | 0.730(0.030) | 0.573(0.108) | 0.809(0.117) | 0.784(0.103) |
4 | 1 | 0.742(0.043) | 0.567(0.101) | 0.819(0.132) | 0.775(0.117) |
0.5 | 2 | 0.776(0.029) | 0.547(0.087) | 0.838(0.074) | 0.836(0.047) |
1 | 2 | 0.769(0.011) | 0.584(0.141) | 0.824(0.117) | 0.818(0.044) |
1.5 | 2 | 0.780(0.047) | 0.572(0.119) | 0.869(0.088) | 0.814(0.071) |
2 | 2 | 0.761(0.034) | 0.603(0.090) | 0.782(0.150) | 0.826(0.038) |
2.5 | 2 | 0.766(0.017) | 0.572(0.081) | 0.820(0.098) | 0.790(0.076) |
3 | 2 | 0.746(0.044) | 0.586(0.114) | 0.777(0.129) | 0.779(0.124) |
Prediction results
We limited the events to those that lasted for at least 30 seconds to investigate the effects of AF duration on predicting episodes of AF with RVR with low activity. As the time to the event increases, the mean prediction performance gradually declines while the variance increases. Using the MAD threshold relative to the whole group, the average AUC for 0.5 min-long prediction intervals is , 1 min-long is and 2 minute-long is (Fig. 6). The 2 minute-long prediction intervals also showed the highest AUC around 0.75 with smallest standard deviations. Prediction results for various gap intervals and prediction intervals are shown in Table 3.
DPFA model interpretation
In addition to the superior performance, another major advantage of our approach is the interpretability of the underlying DPFA models. Indeed the DPFA models could be understood by first calculating the relative frequencies of the states, and then identifying the most prevalent rhythm patterns represented by the states within the DPFA model constructed from each class. This could be achieved via standard techniques by first extracting the normalized leading eigenvector of the transition matrix of the DPFA model, and then taking the average of the relative frequencies across the five cross-validation folds. For example, by comparing the two DPFA models and with signal length (mins) and gap length (mins), this approach yields the following five rhythm patterns as showing the most significant difference between and : $ff (which corresponds to the rhythm pattern , where corresponds to a single window with activity level and ECG type as defined by equation (3, 4, 5, 6)), $ciic, $ici, $ffic, and $cii. Here for ease of presentation we have switched to an alphabet consisting of single-letter symbols, the two alphabets correspond to each other as follows
Old alphabet | |||||||||
New alphabet | a | b | c | d | e | f | g | h | i |
and we use the $ symbol to denote the empty string. Please see Fig. 7 for a plot of the relative frequencies of all the states and the ten most significant rhythm patterns with signal length (mins) and gap length (mins). Please also see Table 4 for the five states with most difference between the relative frequencies in DPFA versus DPFA for various gap intervals and prediction intervals.
Table.4.
Gap interval (min) | Prediction interval (min) | State1 | State2 | State3 | State4 | State5 |
---|---|---|---|---|---|---|
0.5 | 0.5 | $ff | $ciic | $ici | $ffic | $cii |
1 | 0.5 | $i | $ciic | $cii | $icii | $iffff |
1.5 | 0.5 | $iiiciici | $iiiciii | $ciic | $icci | $ciici |
2 | 0.5 | $ff | $cii | $ici | $cifi | $f |
2.5 | 0.5 | $ficific | $ific | $ificic | $iiiciii | $iffi |
3 | 0.5 | $fi | $cii | $ficific | $ciic | $icciic |
3.5 | 0.5 | $ficific | $iccifici | $iiiciicic | $iccii | $icific |
4 | 0.5 | $ff | $cii | $i | $fi | $icc |
4.5 | 0.5 | $cifici | $icif | $cific | $cifi | $fici |
0.5 | 1 | $iicciiiiciii | $iicciiiic | $iicciiiicii | $iicciiiici | $iicciiicii |
1 | 1 | $ificificifici | $ificificific | $iicifii | $iicifiic | $ciicicii |
1.5 | 1 | $ficifici | $cific | $ciici | $cif | $ficificificif |
2 | 1 | $ciic | $cii | $cific | $ciii | $iccii |
2.5 | 1 | $cific | $ciii | $icif | $ciic | $cicici |
3 | 1 | $ificiiic | $ii | $cificif | $ficif | $iccic |
3.5 | 1 | $i | $ic | $f | $if | $iiciici |
4 | 1 | $ificificifici | $icci | $iiccificii | $iiccifii | $ificificific |
0.5 | 2 | $ificificificifici | $iiiiciiiicii | $iiiiciiciicii | $iiiiciiiiciii | $ificific |
1 | 2 | $iiiiciicii | $ificific | $ifiic | $ificificificif | $ificificifici |
1.5 | 2 | $ificificifi | $ificificificific | $ii | $ficifici | $ificificificifici |
2 | 2 | $ici | $icifi | $icif | $ffff | $iccificifi |
2.5 | 2 | $iiccif | $ificificific | $iicc | $iiccic | $iiciiii |
3 | 2 | $iccii | $fici | $ficific | $ficif | $icif |
Discussion
This study in patients with AF showed that our novel DPFA algorithm can predict the onset of AF with RVR associated with low levels of activity with AUC around 0.75 for intervals up to 4.5 min. before the onset of the event. This is the first study to evaluate the performance of prediction algorithms using the ECG along with accelerometer data. Prediction of AF with rapid rates can result in personalized treatment options for prevention and management of AF episodes that are likely to be clinically significant. Our algorithm can help to distinguish between AF with RVR episodes that occur unexpectedly at low levels of activity, thus more likely to be clinically significant.
The machine learning techniques have shown impressive capability to analyze massive amounts of data and are an effective method for classification of arrhythmias using ECG data. Despite being most commonly used in many areas, deep learning algorithms have the drawback that their architecture represents a “Black Box” [8]. Their lack of transparency limits their clinical adoption, makes mechanistic interpretation difficult and reduces their trustworthiness. Further, the disclosure of meaningful details about medical treatment to patients requires the doctors to grasp the fundamental inner workings of the devices they use to some degree [9]. Explainability may also be required to justify the clinical validation of machine learning algorithms in prospective studies and randomized clinical trials [10]. Our DPFA model represents a novel and explainable algorithm. This allows others to validate our algorithm in a diverse population of patients and adopt it clinically if proven to be effective. Previous studies have used the features extracted from the accelerometer data to give an overall impression of patient activity over a period of time. However, these features can not fully capture the variability in daily activity. It also misses the features within the activity signal that can provide important physiological insights about the patient’s condition [11]. Our approach shows that using the one-dimensional accelerometer magnitude from the tri-axial raw accelerometer data can provide important information that can augment the interpretation of the ECG and accelerometer signals beyond the traditional features extracted from the signals.
Most methods for physiological data analysis depend heavily on pre-processing. However, these methods tend to be less effective on noisy data, such as data collected in real-time or in outpatient settings. Therefore, it is desirable to introduce new methods that require minimal pre-processing to analyze such data, thus allowing these insights to be applied to automated clinical decision making. In the previous study, the DPFA method has showed to be particularly useful for the real-world noisy data. Our algorithm does not rely on peak detection, requires minimal pre-processing and leads to good performance in the setting of noisy ECG signal. This provides a significant advantage over existing algorithms when it is necessary to perform rapid analysis in highly noisy environments. Thus DPFA algorithm can be useful for signals captured by portable devices which are prone to noise.
Prediction can only be helpful if it results in specific action that can impact clinical outcomes. In recent years, an increasing number of portable devices have been developed to monitor the physiological signals. Our algorithm provides a possibility for real-time prediction of clinically meaningful arrhythmias using ECG signal together with accelerometer signals with enough lag time for medical interventions.
This is the first algorithm that can identify features within the ECG that are capable of detecting and predicting periods of AF with RVR that are not associated with high activity levels. This allows clinicians to distinguish between clinically significant periods of AF and other periods. This has significant implications for the appropriate treatment of patients with AF. Identifying physiologically significant episodes of AF allows an opportunity to deliver just-in-time treatments that can be tailored for each individual while avoiding the side effects related to daily medications.
Limitations
The first limitation of our study is the sample size. Our algorithm was validated in a small subset of patients with AF. Although the number of patients included in the study was small, there were numerous episodes available for training our algorithm. The validity and generalizability of our algorithm needs to be tested in a more diverse group of patients with AF. We sought to mediate the impact of patient subtypes by leveraging a large number of episodes of AF in our patient population; however, future work incorporating medication status, co-morbidities and other factors would allow for measuring algorithm performance within specific patient subsets while ensuring equal representation in the training dataset.
The aim of our project was to predict AF with rapid ventricular rates and low activity levels. Worsening AF symptom severity is associated with reduced daily activity [12]. The second limitation of the study is a lack of patient-reported symptoms during periods of AF with RVR and low activity. Therefore, it is difficult to ascertain symptom severity during these episodes. Future studies with frequent momentary assessment of symptoms are needed to determine the relationship between symptoms and AF with RVR episodes and low activity.
The third limitation is the interpretability of the activity threshold. In the study, an arbitrary choice of 0.75 quantile MAD based on activity of the entire group is used for classifying activity level. However, participants might have large differences in activity level. To account for these differences, MAD threshold relative to individual participants is also used for analysis. The average for 0.5 min-long prediction intervals is , 1 min-long is and 2 min-long is . The 2 min-long prediction intervals showed most consistent results with around 0.75 and smallest standard deviations. When is less than 1.5 min, the ranges between 0.74-0.80. The is above 0.69 for all prediction intervals. We did not distinguish between different types of activity. It is possible that certain forms of exercise (e.g., stationary cycling) may not be accounted for in the present analysis. The MAD threshold used in the study is a relative MAD threshold instead of an absolute MAD threshold. Studies have investigated different types of activities with their MAD thresholds. However these studies [6] used accelerometer sensors attached to the hip or wrist regions while our sensors are attached to the chest which make our MAD measurements incomparable with these thresholds. There are several studies that investigate activity classification using chest mounted tri-axial accelerometer devices [13–16] with signal processing and machine learning techniques, however they did not state thresholds of activities with MAD metric. We tried to minimize this limitation by using both intra-individual and sample-level activity distributions for the threshold. However, in the future if accelerometer data can be collected with similar sensors and attached to wrists, the prediction results can be better generalized. Both the accelerometer data and ECG data are collected as continuous signals with timestamp when the device is attached to the participant, so the non-wear time information is available and automatically removed, since our study uses synchronized accelerometer and ECG signals. The non-wear time is unlikely to affect our results on prediction of RVR with low activity level. But for future studies it could be helpful to know the reason why the patient removed the device and if it was a consequence of potential events. To this end, we plan to standardize and automate the annotation process by intelligently mining the training data. We are hopeful that such efforts will continue to improve of the performance of the algorithm and further expand the scope of our approach.
Conclusion
In this paper, a DPFA based method is performed for predicting the onset of AF with RVR associated with low levels of activity up to 4.5 min prior to the onset of the events. The proposed algorithm takes input from two different types of synchronized physiological signals, constructs the state space and the underlying transition probabilities directly from the data input. The model achieved an overall AUC around 0.7–0.8 depending on the lengths and gap sizes of the prediction intervals. Although the database is relatively small in terms of number of participants, each participant has up to 3 weeks of continuous synchronized ECG and accelerometer data. The performance of the proposed DPFA algorithm does not depend on pre-processing the data. With a larger dataset and additional validation, our algorithm could enable the development of just-in-time intervention alert system on wearable portable devices that could reduce the morbidity and mortality associated with AF and other similar arrhythmias.
Acknowledgements
The development of the algorithm in this paper was funded by the Collaborative Safety Research Center of Toyota Motor North America. The authors would like to thank Amanda Gomez for her help on preparation and submission of the manuscript.
Abbreviations
- AF
Atrial fibrillation
- AHRE
Atrial high-rate episode
- AUC
Area under the ROC curve
- BPM
Beats per minute
- DPFA
Deterministic probabilistic finite-state automata
- ECG
Electrocardiogram
- HR
Heart rate
- IQR
Interquartile range
- MAD
Mean amplitude deviation
- RVR
Rapid and irregular ventricular rate
- SVT
Supraventricular tachycardia
Authors’ contributions
ZL was involved with conception, design, acquisition and analysis, interpretation of data, drafting and revision of the work. HG was involved with conception, design, acquisition and analysis; interpretation of data, drafting and revision of the work. KW was involved with conception, design, acquisition and analysis, interpretation of data, drafting and revision of the work. SL-P was involved with conception, design, acquisition and analysis, interpretation of data, drafting and revision of the work. HO was involved with conception, design, acquisition and analysis, interpretation of data, drafting and revision of the work. DC was involved with conception, design, acquisition and analysis, interpretation of data, drafting and revision of the work. PG was involved with conception, design, acquisition and analysis, interpretation of data, drafting and revision of the work. JG was involved with conception, design, acquisition and analysis, interpretation of data, drafting and revision of the work. KN was involved with conception, design, acquisition and analysis, interpretation of data, drafting and revision of the work. BN was involved with conception, design, acquisition and analysis, interpretation of data, drafting and revision of the work. All authors read and approved the final manuscript.
Funding
The funding source for this study includes, K23 Career Development Grant (1K23HL135397).
Availability of data and materials
The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.
Declarations
Ethics approval and consent to participate
University of Michigan IRB approved the protocol and written informed consent was obtained. All methods were carried out in accordance with relevant guidelines and regulations.
Consent for publication
Not applicable.
Competing interests
Dr.Ghanbari has research support from National Institutes of Health and Toyota Corporation. Dr.Ghanbari serves as a consultant for Preventice and Verily Inc.
Footnotes
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.January CT, Wann LS, Alpert JS, Calkins H, Cigarroa JE, Cleveland JC, Conti JB, Ellinor PT, Ezekowitz MD, Field ME, et al. 2014 aha/acc/hrs guideline for the management of patients with atrial fibrillation: executive summary: a report of the American college of cardiology/American heart association task force on practice guidelines and the heart rhythm society. J Am Coll Cardiol. 2014;64(21):2246–2280. doi: 10.1016/j.jacc.2014.03.021. [DOI] [PubMed] [Google Scholar]
- 2.Skinner NS, Jr, Mitchell JH, Wallace AG, Sarnoff SJ. Hemodynamic consequences of atrial fibrillation at constant ventricular rates. Am J Med. 1964;36(3):342–350. doi: 10.1016/0002-9343(64)90160-3. [DOI] [PubMed] [Google Scholar]
- 3.Kochiadakis G, Skalidis E, Kalebubas M, Igoumenidis N, Chrysostomakis S, Kanoupakis E, Simantirakis E, Vardas P. Effect of acute atrial fibrillation on phasic coronary blood flow pattern and flow reserve in humans. Eur Heart J. 2002;23(9):734–741. doi: 10.1053/euhj.2001.2894. [DOI] [PubMed] [Google Scholar]
- 4.Godfrey A, Conway R, Meagher D, ÓLaighin G. Direct measurement of human movement by accelerometry. Med Eng Phys. 2008;30(10):1364–1386. doi: 10.1016/j.medengphy.2008.09.005. [DOI] [PubMed] [Google Scholar]
- 5.Teplitzky BA, McRoberts M, Ghanbari H. Deep learning for comprehensive ECG annotation. Heart Rhythm. 2020;17(5):881–888. doi: 10.1016/j.hrthm.2020.02.015. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Bakrania K, Yates T, Rowlands AV, Esliger DW, Bunnewell S, Sanders J, Davies M, Khunti K, Edwardson CL. Intensity thresholds on raw acceleration data: Euclidean norm minus one (enmo) and mean amplitude deviation (mad) approaches. PLoS ONE. 2016;11(10):0164045. doi: 10.1371/journal.pone.0164045. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Li Z, Derksen H, Gryak J, Jiang C, Gao Z, Zhang W, Ghanbari H, Gunaratne P, Najarian K. Prediction of cardiac arrhythmia using deterministic probabilistic finite-state automata. Biomed Signal Processing Control. 2021;63:102200. doi: 10.1016/j.bspc.2020.102200. [DOI] [Google Scholar]
- 8.Topol EJ. High-performance medicine: the convergence of human and artificial intelligence. Nat Med. 2019;25(1):44–56. doi: 10.1038/s41591-018-0300-7. [DOI] [PubMed] [Google Scholar]
- 9.Nagendran M, Chen Y, Lovejoy CA, Gordon AC, Komorowski M, Harvey H, Topol EJ, Ioannidis JP, Collins GS, Maruthappu M. Artificial intelligence versus clinicians: systematic review of design, reporting standards, and claims of deep learning studies. BMJ 2020; 368. [DOI] [PMC free article] [PubMed]
- 10.Doshi-Velez F, Kim B. Towards a rigorous science of interpretable machine learning. arXiv preprint arXiv:1702.08608 (2017)
- 11.Yang C-C, Hsu Y-L. A review of accelerometry-based wearable motion detectors for physical activity monitoring. Sensors. 2010;10(8):7772–7788. doi: 10.3390/s100807772. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Semaan S, Dewland TA, Tison GH, Nah G, Vittinghoff E, Pletcher MJ, Olgin JE, Marcus GM. Physical activity and atrial fibrillation: data from wearable fitness trackers. Heart Rhythm. 2020;17(5):842–846. doi: 10.1016/j.hrthm.2020.02.013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Godfrey A, Bourke A, Olaighin G, Van De Ven P, Nelson J. Activity classification using a single chest mounted tri-axial accelerometer. Med Eng Phys. 2011;33(9):1127–1135. doi: 10.1016/j.medengphy.2011.05.002. [DOI] [PubMed] [Google Scholar]
- 14.Gjoreski H, Rashkovska A, Kozina S, Lustrek M, Gams M. Telehealth using ECG sensor and accelerometer. In: 2014 37th international convention on information and communication technology, electronics and microelectronics (MIPRO). IEEE, 2014; pp. 270–274
- 15.Purwar A, Jeong DU, Chung WY. Activity monitoring from real-time triaxial accelerometer data using sensor network. In: 2007 International conference on control, automation and systems. IEEE, 2007; pp. 2402–2406
- 16.Gao L, Bourke A, Nelson J. Evaluation of accelerometer based multi-sensor versus single-sensor activity recognition systems. Med Eng Phys. 2014;36(6):779–785. doi: 10.1016/j.medengphy.2014.02.012. [DOI] [PubMed] [Google Scholar]