Published in final edited form as: IEEE Trans Med Imaging. 2021 Nov 30;40(12):3389–3399. doi: 10.1109/TMI.2021.3086331

TABLE I.

The Venc, Intensity Based Signal-to-Noise Ratio (SNRI), Number of Wrapped Voxels (NW), Velocity-to-Noise Ratio (VNR), Mean WSS Magnitude, and Relative WSS Magnitude Error (ϵWSS) for the Each in vitro Poiseuille Flow Dataset With CWLS and 4D Lap Unwrapping

DV acquisitions A B C A B C
venc (cm/s) 4 6 8 8 12 16
TE (ms) 7.47 6.47 5.87 7.47 6.47 5.87
TR (ms) 10.2 9.2 8.6 10.2 9.2 8.6
SNRI 60.9 54.1 47.9 60.9 54.1 47.9
Nw 41919 32819 24128 23434 3925 1
Nw CWLS 1 0 0 0 0 0
4D Lap 104 2 20 38 4 10
VNR CWLS 33.2 38.4 28.1 16.0 18.5 14.4
4D Lap 23.7 26.4 17.5 10.7 13.3 9.4
VNR improve (%) 40 46 61 50 39 53
mean WSS (Pa) CWLS 0.17 0.16 0.17 0.20 0.18 0.20
4D Lap 0.20 0.17 0.19 0.23 0.20 0.22
relative ϵWSS CWLS 0.45 0.37 0.42 0.68 0.51 0.65
4D Lap 0.70 0.56 0.78 1.38 1.02 1.50
ϵWSS reduction (%) 56 53 85 105 102 130