Table 2. Sensitivity, specificity and negative predictive value of R-time to detect DOAC concentrations above thresholds mentioned.

DOAC Threshold R-time Sensitivity Specificity LR+ LR− NPV
≥(ng/mL) ≥(min) % % %
Dabigatran
30 2.6 100 92 (82–100) 13 0 100
50 3.1 94 (88–100) 83 (70–95) 5.5 0.07 91 (81–100)
100 3.4 100 82 (72–93) 5.6 0 100
Rivaroxaban
30 1.7 98 (94–100) 86 (73–99) 6.9 0.02 96 (88–100)
50 2.1 95 (89–100) 80 (67–93) 4.8 0.06 93 (84–100)
100 2.6 96 (78.9–99.9) 85 (72.4–93.3) 5.1 0.05 98 (93–100)
Apixaban
30 c
50 1.7 100 96 (88–100) 25 0 100
100 2.2 98 (93–100) 81 (67–95) 5 0.03 96 (89–100)

Abbreviations: AFXa, antifactor-Xa; AUC, area under the curve; DOAC, direct oral anticoagulant; DTI, direct thrombin inhibitor; LR + , positive likelihood ratio; LR − , negative likelihood ratio; NPV, negative predictive value; R, reaction time.

Note: The reference range of R-time for the FXa channel is 0.6–1.5 minutes, and for DTI channel is 1.6–2.5 minutes.

a

Indicates using the DTI channel.

b

Indicates using the AFXa channel.

c

Due to only single acquired concentration between 30 and 49 ng/mL, no calculations were performed.