. 2021 Sep;32(3):137–147. doi: 10.7171/jbt.21-32-03-009

TABLE 4.

Samples identified as potentially positive for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) by reverse-transcription loop-mediated isothermal amplification (RT-LAMP during point-of-care (POC) testing


RT-LAMP-positive sample

qRT-PCR viral load, copies/μL
POC1 8.53
POC2 2.15 × 104
POC3 Negative
POC4 Negative
POC5 Negative
POC6 Negative
POC7 3.62 × 105
POC8 Negative
POC9 n/aa
POC10 2.12 × 103
POC11 Negative
POC12 1.04 × 103
POC13 2.06 × 102
POC14 Negative
POC15 52.8
POC16 6.02 × 102
POC17 87.3
POC18 1.17 × 103
POC19 Negative
POC20 1.38 × 102
POC21 4.07 × 102

Abbreviations: n/a, sample not available; qRT-PCR, quantitative reverse-transcription polymerase chain reaction.

a

Volunteer did not consent to follow-up testing.