Skip to main content
. 2021 Oct 9;41(1):175–179. doi: 10.1007/s10096-021-04350-y

Table 2.

Categorical agreement between the ReaScan and recomBead CXCL13 assays, with recomBead as reference (n = 209)

recomBead1 ReaScan2 Agreement (%)
Negative Grey zone Positive
Negative 140 137 2 1 98
Grey zone 10 10 0 0 0
Positive 59 8 7 44 75

1recomBead CXCL13 negative; CXCL13 < 190 pg/mL, grey zone; CXCL13 191–300 pg/mL, positive; CXCL13 > 300 pg/mL. 2ReaScan CXCL13 negative; CXCL13 < 250 pg/mL, grey zone; CXCL13 250–500 pg/mL, positive; CXCL13 > 500 pg/mL