Table 2. Relevant criteria for the synchrotron radiation spectra in 3 T with different filtration modes.
An energy threshold of 20–500 keV is applied to the data set with a bin size of 0.1 keV. TD stands for transmission detector.
| Criterion | Mode | Without TD | With 10 µm TD | With 375 µm TD |
|---|---|---|---|---|
| Maximum energy (keV) | Al/Al | 39.9 ± 0.1 | 39.9 ± 0.1 | 41.9 ± 0.1 |
| Mean energy (keV) | Al/Al | 35.2 ± 0.1 | 35.1 ± 0.1 | 35.4 ± 0.1 |
| Maximum energy (keV) | Cu/Al | 67.9 ± 0.1 | 67.9 ± 0.1 | 68.9 ± 0.1 |
| Mean energy (keV) | Cu/Al | 61.9 ± 0.1 | 61.9 ± 0.1 | 62.0 ± 0.1 |
| Maximum energy (keV) | Cu/Cu | 80.9 ± 0.1 | 80.9 ± 0.1 | 79.9 ± 0.1 |
| Mean energy (keV) | Cu/Cu | 71.1 ± 0.1 | 71.2 ± 0.1 | 71.3 ± 0.1 |