. 2022 Jan 1;29(Pt 1):125–137. doi: 10.1107/S1600577521011140

Table 6. FWHM, peak/valley dose and PVDR for microbeams at 5, 10 and 75 mm depth with and without the 375 µm transmission detector (TD) in place upstream.

Measurements are performed in the water phantom using the PTW microDiamond in 3 T mode with Al/Al filtration.

  5 mm depth 10 mm depth 75 mm depth
Parameter With TD Without TD Difference (%) With TD Without TD Difference (%) With TD Without TD Difference (%)
FWHM (µm) 45.5 46.8 2.78% 46.1 45.8 −0.66% 48.4 47.9 −1.04%
Peak dose (Gray) 412.89 438.45 5.83 366.71 384.50 4.63 76.16 79.87 4.65
Valley dose (Gray) 22.51 24.03 6.32 22.59 24.09 6.21 6.94 7.01 1.03
PVDR 18.34 18.25 −0.53 16.23 15.96 −1.69 10.98 11.39 3.66