Table 2.

Average quality and depth-of-coverage metrics across validation samples

Run metrics*
Sample coverage**
ISP density, % yield total reads (M) useable reads % Polyclonal mapped reads % on target mean depth uniformity, %
FNAB 94 22.6 64% 31 1,380,181 96 2,430 93
FFPE 93 18.7 53% 38 1,326,909 96 2,288 93
Acceptable >70 10–30 >30% <50 >200,000 >80 ≥500X ≥80

FFPE, formalin-fixed paraffin-embedded; FNAB, fine-needle aspiration biopsy.

*

Run metrics were averaged from 9 separate runs for FNAB samples and 8 separate runs for FFPE samples.

**

Sample coverage data were averaged from the 68 FNAB samples and 37 FFPE samples used in the validation study.