Detection and measurement of midline shift using portable MRI.

Sensitivity Specificity Dichotomous pMRI vs. SOC (κ) Continuous pMRI vs. SOC (ICC)
All SW Versions 0.93 0.96 0.87 0.94
RC3/RC4a 0.80 0.95 0.75 0.73
RC5/RC6b 1.00 0.96 0.90 0.97
RC7/RC8c 1.00 1.00 1.00 1.00

SW, software; κ, kappa statistic; ICC, intraclass correlation coefficient; pMRI, portable MRI; SOC, standard-of-care imaging.

aPatients scanned using pMRI software versions RC3 and RC4.

bPatients scanned using pMRI software versions RC5 and RC6.

cPatients scanned using pMRI software versions RC7 and RC8.