Table 3.
Category | Recommendation | Status |
User interface | Add label to console screen buttons | Updated |
Improve drapes and drape connectors | Updated | |
Increase light intensity of orientation arrow | Updated | |
User manual | Add direct references to troubleshooting in the user manual | Planned for future version |
Expand and review the index in the instrument and accessories manual so that readers find terms more easily | Planned for future version | |
Emphasize how to change from trainer to surgical mode on console | Implemented* | |
Training | Emphasize drape attachment technique | Implemented |
Emphasize optimal BSU positioning | Implemented | |
Emphasize cart orientation | Implemented | |
More time for bedside team at the console to improve their understanding of using the hand controllers and menu navigation | Implemented | |
More time for surgeons at the bedside to improve their knowledge of the arm mode map and general use of the BSUs including sleep button, brake button orientation | Implemented | |
Emphasize docking features can be stowed under console arms | Implemented | |
System issues | Improve console’s brake | Updated |
Improve camera connection | Updated | |
HUD internal component upgraded | Updated |
*A new chapter on Versius Trainer will be added to the user manual.
BSU, bedside unit; HUD, head-up display.